• In the following situations it is recommended that three temperatures in the same ear be taken and
the highest one taken as the reading:
1) New born infants in the first 100 days.
2) Children under three years of age with a compromised immune system and for whom the presence or
absence of fever is critical.
3) When the user is learning how to use the IR thermometer for the first time until he/she has familiar-
ized himself/herself with the instrument and obtains consistent readings.
7.
Changing from Fahrenheit to Celsius and vice-versa
The Ear Thermometer TH 1DB1-1 can display temperature measurements in either Fahrenheit or Celsius.
To switch the display between °C and °F, simply turn OFF the unit , press and hold the Start button for 5
seconds; When you stop pressing the Start button after 5 seconds, the current measurement scale (°C or
°F icon) will be flashed on the display. Toggle the measurement scale between °C and °F by pressing the
start button again. When the measurement scale has been chosen, wait for 5 seconds and the unit will
enter the ready for measuring mode automatically.
8.
How to recall 12 readings in Memory Mode
The Probe Cover Free Thermometer TH 1DB1-1 can recall the last 12 readings. To recall please follow
sequence.
Pressing and releasing the START button after the last 12 readings have been recalled will resume the
above sequence from reading 1.
6
LCD display
Display Meaning
Description
Recall Mode
Press the START button to enter Recall Mode
when power is off. The memory icon «
M
»
flashes.
Reading 1 (the last
reading)
Press and release the START button to recall
the last reading. Display 1 alone with memory
icon.
Reading 2 (the second
latestreading)
Press and release the START button to recall
the second latest reading
Reading 12 (the last
reading)
Press and release the START button consecu-
tively to recall readings in succession, up to the
last 12 readings.
Содержание TH 1DB1-1