background image

The 2 wire, 1000 ohm RTD sensor must be connected
across terminals #1 AND #2 of the terminal block.

Calibration Procedure for the 2 Wire RTD:

RTD devices are precision resistors whose resistance
value varies with temperature. The DE20 control mea-
sures the RTD resistance and compares that mea-
surement with a “standard” set of values stored in the
control memory. To restore, update or verify that this
“standard” set of values is correct, do the following:

Equipment needed:

1.

Two precision resistors (tol/- 0.1%
or better) with a fixed value equal to the
nominal value of the RTD (i.e. 1000 ohms).

2.

A suitable jumper cable to facilitate
changing input resistance.

WARNING!

Calibration procedures require the removal of the
rear cover of the control. It also requires that power
is ON, exposing the technician to potentially lethal
voltages. Exercise EXTREME CARE and wear tested
electrician’s gloves whenever power is on.

Step 1:

Turn 

OFF

 all power.

Step 2:

Remove rear cover.

Step 3:

Remove RTD sensor.

Step 4:

Install the precision resistors in place of the

RTD sensor, as shown.

-6-

2 Wire RTD Sensor Calibration

Step 5:

Install the jumper cable between the loose

end of one of the resistors and the fixed end of the
other resistor to establish an input value of a single
resistor (i.e. 1000 ohms), as shown.

Step 6:

Carefully restore power to the control, taking

precautions not to make contact with any exposed
voltage sources.

Step 7:

Press the decrease “

È

” key and increase

Ç

” buttons simultaneously and hold for approximately

6 seconds.  The display will indicate “AC.0”.  While
the “0” is flashing, use the increase key to change
this to “22”. Press the “SET” key. The control display
will read “

CAL1

”.

Step 8:

Press and hold the “

SET

” key for 1 second.

The display will read “

HoLd

”. Wait for the display to

change to “

CAL2

”.

Step 9:

Proceed with caution to avoid SHOCK haz-

ard. Remove and relocate one end of the jumper cable
to the loose end of the second precision resistor for
the second resistance value (i.e. 2000 ohms), as
shown.

Step 10:

 Press and hold the “

SET

” key for 1 second.

The display will read “

HoLd

”. WAIT for the display to

reset. After resetting, the display should indicate the
approximate temperature value for the connected pre-
cision resistors (i.e. 511ºF or 266ºC).

Step 11: 

Turn OFF power and remove the precision

resistors. Reinstall the RTD sensor and the rear cover
of the controller. Return the calibrated control to service.

M-33-01-02  02/17/11

Make sure that the DIP switch settings are OFF, ON,
OFF (see page 13). The voltage signal must be con-
nected across terminals #1 and #2 of the Adder Board
(Item Number 5415). Terminal #1 is common (nega-
tive), and terminal #2 is the signal connection (positive).

Note: Polarity must be observed.

WARNING!

Calibration procedures require the removal of the
rear cover of the control. It also requires that power
is ON, exposing the technician to potentially lethal
voltages. Exercise EXTREME CARE and wear tested
electrician’s gloves whenever power is on.

Calibration procedure for voltage signal:

Step 1:

Turn 

OFF

 all power.

Step 2:

Remove rear cover.

Step 3:

Remove sensor.

Step 4:

Install the voltage calibrator to terminal #1

and #2.

Step 5:

CAREFULLY restore power to the controller,

ensuring that you do not come in contact with any
exposed voltage.

Step 6:

Press the decrease “

È

” key and increase

Ç

” buttons simultaneously and hold for approximately

6 seconds.  The display will indicate “AC.0”.  While
the “0” is flashing, use the increase key to change
this to “22”.  Press the “SET” key.  The control display
will read “

CAL1

”.

Step 7:

Adjust the power supply to 1.0 volts.

Step 8:

Press and hold the “

SET

” key for one sec-

ond. The display will read “

HoLd

”. Wait for the display

to change to “

CAL2

”.

Step 9:

Adjust the power supply to 10.0 volts.

Step 10:

 Press and hold the “

SET

” key for one sec-

ond. The display will read “

HoLd

”. Wait for the display

to reset and display 10.0.

Step 11:

 Turn OFF power to the control and remove

the power supply. Reinstall the voltage input and
the rear cover of the control. Return the calibrated
control to service.

The DE20 control measures the DC current and com-
pares that measurement with a “standard” set of val-
ues in the control memory. To restore, update or
merely verify that this “standard” set of values is cor-
rect, do the following:

Check that the DIP switches are set to OFF, OFF, ON
(see page 13).

-9-

Current Input Calibration

Equipment needed:

1.

A precision, NIST traceable, 0-20 mA DC
current calibrator, OR

2.

A precision, NIST traceable, digital
ammeter or DMM and:
a.

a regulated linear DC power
supply with an adjustable
0-10 volt or better output, and

b.

a 400 ohm, 0.1% or better tolerance,
precision resistor.

WARNING!

Calibration procedures require the removal of the
rear cover of the control. It also requires that power
is ON, exposing the technician to potentially lethal
voltages. Exercise EXTREME CARE and wear tested
electrician’s gloves whenever power is on.

Calibration procedure for a current input:

Step 1:

Turn 

OFF

 all power.

Step 2:

Remove rear cover.

Step 3:

Remove input leads.

Step 4:

Install the 0-20 mA DC calibrator or the power

supply, resistor and ammeter in series with terminal
#1 and #2.

Step 5:

CAREFULLY restore power to the controller,

ensuring that you do not come in contact with any
exposed voltage.

Step 6:

Press the decrease “

È

” key and increase

Ç

” buttons simultaneously and hold for approximately

6 seconds.  The display will indicate “AC.0”.  While
the “0” is flashing, use the increase key to change
this to “22”.  Press the “SET” key.  The control display
will read “

CAL1

”.

Step 7:

Adjust the calibrator or power supply to 5.0

mA.

Step 8:

Press and hold the “

SET

” key for one sec-

ond. The display will read “

HoLd

”. Wait for the display

to change to “

CAL2

”.

Step 9:

Adjust the power supply to 20.0 mA.

Step 10: 

Press and hold the “

SET

” key for one sec-

ond. The display will read “

HoLd

”. Wait for the display

to reset and display 20.0.

Step 11: 

Turn OFF power to the control and remove

the power supply. Reinstall the voltage input and the
rear cover of the control. Return the calibrated control
to service.

M-33-01-02  02/17/11

U

U

U

U

U

!

U

U

U

U

U

!

U

U

U

U

U

!

Содержание DE20

Страница 1: ...stance Signal Calibration 7 3 Wire RTD Calibration 7 Thermocouple Calibration 8 Voltage Signal Calibration 8 Current Input Calibration 9 Frequency Signal Calibration 10 Configuration Setup 10 Main Men...

Страница 2: ...0 to 1000 F 40 to 538 C F or C field selectable Set Point Range Selectable throughout the input range De energize control output No sensor short protection with Thermocouple sensor Accuracy 0 25 span...

Страница 3: ...Using the wiring dia gram supplied with the controller install the required input and output wires Use National Electric Code and local codes for determining wire sizing insula tion terminations etc 3...

Страница 4: ...Sensor Number represents degrees Celsius or Fahrenheit as determined by the F5 setting Voltage Sensor Number represents tenths of a volt 0 1 VDC Current Sensor Number represents hundredths of milliam...

Страница 5: ...higher than the set point setting on the control During the alarm condition the control relay will de energize the set point relay and flash AAA on the display After enabling this feature you may vie...

Страница 6: ...ugh the list of settings to those needing modification 10 Configuration To adjust a setting while in the Configuration Mode use the increase decrease buttons to bring the particular setting into the v...

Страница 7: ...trician s gloves whenever power is on Calibration procedure for voltage signal Step 1 Turn OFF all power Step 2 Remove rear cover Step 3 Remove sensor Step 4 Install the voltage calibrator to terminal...

Страница 8: ...standard set of values is cor rect do the following THERMOCOUPLE NEGATIVE POSITIVE 7 Optional Item Number 5415 board needed The configuration and calibration procedures used for measuring pure resista...

Отзывы: