Instructions for use
PRIMEDIC
™
Defi-Monitor ECO 1 / DM 1 / DM 3
82
19499 / 05.03
Attention:
Only use the original-
PRIMEDIC
™
patient cable. This patient
cable has an integrated defibrillation protection protecting the
ECG monitor from dangerous high voltage. The use of
another patient cable may cause the destruction of the ECG
unit and may be hazardous to the user.
Attention:
The black ECG electrode (N) always has to be connected, to
guarantee the function of the identification of electrodes
open.
6. Operation
of
SpO
2
monitor (only model
DM 3
)
6.1 Connection of SpO
2
sensor
The
PRIMEDIC
™
Defi-Monitor
with integrated NELLCOR
®
pulsoximeter module is provided with a blue input socket
(16) at the upper left of the casing for the SpO
2
patient cable.
Proceed as follows:
1.
Insert into slot (16) the blue plug (17) of the patient cable
(about 1,4 m).
2.
Plug the NELLCOR
®
sensor into the other end of the SpO
2
patient cable until the traction relief engages.
3.
Test the perfect mechanical locking of the plug-in-
connection.
To dismount the NELLCOR
®
sensor from the SpO
2
patient
cable, push the release of the plug-in-connection and take out
the sensor without applying force.
If the NELLCOR
®
sensor is positioned on a patient, after
about 5 seconds of measuring time the value in % and the
pulse curve are shown on the monitor. A rhythmically
oscillating signal bar at the right fringe of the pulse curve
indicates when the module recognizes an evaluable oxygen
saturation value.
Note:
Do not kink sensor or patient cable to avoid defects or
malfunctions.