Instructions for use
PRIMEDIC
™
Defi-Monitor ECO 1 / DM 1 / DM 3
80
19499 / 05.03
The value indicated on the monitor (e. g.
♥
ALARM 30 / 90,
SpO
2
90 / 100) means that a pulsating acoustic alarm will
be released when exceeding the heartrate of 90 beats/
minute or remaining under 30 beats / minute. You can
acknowledge (switch off) the alarm by pressing ALARM (28)
shortly. The acoustic alarm is switched off for about 1 minute
whereas the alarm limits remain. To change the alarm limits
or to switch off the alarm press ALARM (28) until the desired
selection (e. g.
♥
ALARM — — —) is shown on the monitor.
After the defibrillation the heart alarms are set
OFF
automatically.
Note:
Test possibility for alarm circuit.
Select alarm limits with no patient connected. The optical and
acoustical alarm has to be released immediately. After
acknowledging ALARM (press button 28) the acoustical
alarm is suppressed for about 1 minute. The optical alarm has
to remain active (intermittent alarm signal).
5.5 Contrast-Settings (only for ECO 1)
Five different contrast settings are available. Depending on
the ambient temperature or brightness it may be useful to
change the contrast for high-contrast monitoring. The most
suitable contrast step can be determined by repeated
actuation of CONTRAST (29). After switching off the
PRIMEDIC
™
Defi - Monitor
the current contrast setting is
saved and active again when switching on.
5.6 Filter
The monitoring has the possibility to use high-quality line filter
with 50 or 60 Hz. This can be adjusted in set-up menu (see
section 3.2). It is recommendable to activate the 50 Hz filter.