Instructions for use
PRIMEDIC
™
Defi-Monitor ECO 1 / DM 1 / DM 3
19499 / 05.03
79
5.2 Setting of signal amplification
Five signal amplification steps are available:
1. 0.5
cm/mV
2. 0.8
cm/mV
3. 1.0
cm/mV
4. 1.5
cm/mV
5 2.0
cm/mV
Amplification step 1 cm/mV is automatically active after
switching on the
PRIMEDIC
™
Defi-Monitor
. Press SIGNAL
AMPLIFICATION (23) to switch between the amplification
steps in ring counting method. The 1mV reference is
constantly shown on the left of the monitor border. After
switching to a different value the reference marking is shown
for about 5 seconds in lower status line.
5.3 Volume of systole beep
Three volume steps are available for the systole beep: off, low,
high.
After switching on the volume step
off
is automatically
active. The volume symbol is shown on the monitor.
Press VOLUME (24) to switch between the three volume
steps. Each depression on key (24) causes the continuation
by one step.
5.4 Activation of heartrate alarms
Four steps are available:
1.
— — —
2.
♥
ALARM 1
30 / 90
3.
♥
ALARM 2
40 / 130
4.
♥
ALARM 3
50 / 180
The above mentioned alarm limits are preset at delivery. The
values of the second and fourth alarm step can be changed
by means of the set-up menu (see section 3.2).
The alarm is not active after switching on the
PRIMEDIC
™
Defi-Monitor
which is indicated on the monitor. Press
ALARM (28) for activation. Every depression of this key
causes the continuation by one alarm step.