Troubleshooting Tips (Including FAULT & ERROR Messages)
Symptom
Check/Action
No display at all
Check power at power connector
Not able to change setup or
programming,
Locd
is displayed
Meter is password-protected, enter correct six-digit password to
unlock
Meter displays error message
during calibration (
Error
)
Check:
1. Signal connections
2. Input selected in
Setup
menu
3. Minimum input span requirements
Meter displays
1.
999999
2.
-99999
3.
fault
Check:
1. Input selected in
Setup
menu
2. Signal at Signal connector
3. Input exceeds range selected
Display is unstable
Check:
1. Input signal stability and value
2. Display scaling vs. input signal
3. Filter and bypass values (increase)
4. Increase Rounding value
Display response is too slow
Check filter and bypass values
Display reading is not accurate
Check:
1. Signal input conditioner selected: Linear, square root, etc.
2. Scaling or calibration
Display does not respond to
input changes, reading a fixed
number
Check:
1. Display assignment, it might be displaying max, min, or set point.
Display alternates between
1.
Hi
and a number
2.
Lo
and a number
Press Menu to exit max/min display readings.
Relay operation is reversed
Check:
1. Fail-safe in
Setup
menu
2. Wiring of relay contacts
Relay and status LED do not
respond to signal
Check:
1. Relay action in
Setup
menu
2. Set and reset points
Flashing relay status LEDs
Relays in manual control mode or relay interlock switches opened.
Meter not communicating with
application programs
Check:
1. Serial adapter and cable
2. Serial settings
3. Meter address and baud rate
If the display locks up or the
meter does not respond at all
Cycle the power to reboot the
microprocessor.
Other symptoms not described
above
Call Technical Support for
assistance.