PD6100 Strain Gauge, Load Cell, & mV Meter Instruction Manual
49
Tips:
Low and high input signals can be any valid values within the range of the meter.
Observe minimum input span requirements between input 1 and input 2.
Low input should be less than high input signal.
Error Message (
error
)
An error message indicates that the calibration or scaling process was not successful.
After the error message is displayed, the meter reverts to input 2 during calibration or scaling and to input
1 during internal calibration, allowing the appropriate input signal to be applied or programmed.
The error message might be caused by any of the following conditions:
1. Input signal is not connected to the proper terminals, or it is connected backwards.
2. Wrong signal selection in
Setup
menu.
3. Minimum input span requirements not maintained.
Minimum Input Span
The minimum input span is the minimum
difference between input 1 and input 2 signals required to complete
the calibration or scaling of the meter.
Input Range
Input 1 & Input 2 Span
15 mV
0.2 mV
25 mV, 30 mV
0.4 mV
150 mV
2.0 mV
250 mV, 300 mV
4.0 mV