PD8-6262 Explosion-Proof Dual Analog Input Flow Rate/Totalizer
Instruction Manual
52
Advanced Features Menu
To simplify the setup process, functions not needed
for most applications are located in the
Advanced
Features
menu.
Press and hold the Menu button for three seconds to
access the advanced features of the totalizer.
Advanced Features Menu &
Display Messages
The following table shows the functions and
messages of the
Advanced Features
menu in the
order they appear in the menu.
Advanced Features Menu & Display Messages
Display
Parameter
Action/Setting
filtEr
Filter
Set noise filter value
Ch-A
Channel A
Set filter value for channel
A
Ch-b
Channel B
Set filter value for channel
B
bypASs
Bypass
Set filter bypass value
Ch-A
Channel A
Set filter bypass value for
channel A
Ch-b
Channel B
Set filter bypass value for
channel B
Round
Round
Set the rounding value for
display variables
seriAl
Serial
Set serial communication
parameters
SlaVId
Slave ID
Set slave ID or meter
address
Baud
Baud rate
Select baud rate
Tr dLY
Transmit delay
Set transmit delay for serial
communication
Parity
Parity
Select parity
Even, Odd, or None with 1
or 2 stop bits
t-byt
Time byte
Set byte-to-byte timeout
SelEct
Select
Enter the Select menu
(function, cutoff, out)
Functn
Input signal
conditioning
Select linear, square root,
programmable exponent, or
round horizontal tank
function
Ch-A
Channel A
Select menu for channel A
Advanced Features Menu & Display Messages
Display
Parameter
Action/Setting
Ch-b
Channel B
Select menu for channel B
Linear
Linear
Set meter for linear function
and select number of
linearization points
No pts
Number of
points
Set the number of
linearization points
(default: 2)
Square
Square root
Set meter for square root
extraction
Prog E
Programmable
exponent
Set meter for
programmable exponent
and enter exponent value
nmath
Math
Enter the setup menu for
channel C math functions
Sunm
Sum
C = (A+B+P)*F
Dif
Difference
C = (A-B+P)*F
difabs
Absolute
difference
C = ((Absolute value of (A-
B))+P)*F
AVg
Average
C = (((A+B)/2)+P)*F
nmulti
Multiplication
C = ((A*B)+P)*F
DiVide
Divide
C = ((A/B)+P)*F
Hi-ab
Max of A or B
C = ((High value of channel
A or B)+P)*F
Lo-ab
Min of A or B
C = ((Low value of channel
A or B)+P)*F
Drauw
Draw
C = ((A/B)-1)*F
uw Avg
Weighted avg.
C = ((B-A)*F)+A
ratio
Ratio
C = (A/B)*F
Ratio2
Ratio 2
C = ((B-A)/A)+P)*F
Concen
Concentration
C = (A/(A+B))*F
Sunm t
Sum total
C = (tA+tB+P)*F
Sunmgt
Sum grand total
C = (GtA+GtB+P)*F
Dif t
Difference of
total
C = (tA-tB+P)*F
Dif gt
Diff. of grand
total
C = (GtA-GtB+P)*F
Tratio
Total ratio
C = (tA/tB)*F
t-rat2
Total ratio 2
C = ((tB-tA)/tA)*F
T pct
Total percent
C = (tA/(tA+tB))*100
Const
Constant
Enter math equation
constants
adder
Adder
Addition constant used in
channel C math
calculations (P)
factor
Factor
Multiplication constant used
in channel C math
calculations (F)
CutofF
Cutoff
Set low-flow cutoff
Ch-A
Channel A
Set low-flow cutoff for
Channel A