PD6928 Ex-Proof & I.S. Loop-Powered Flow Rate/Totalizer
Instruction Manual
69
Meter
Operation
The meter is equipped with four CapTouch through-
glass buttons. These buttons allow meter operation
without removing the cover and exposing the
electronics in a hazardous area.
Three CapTouch buttons labeled F1, F2, and F3 can
be programmed as function keys to perform a variety
of meter functions simply by pressing on the glass
over the button. These include resetting the total,
resetting the meter’s relays or open collectors,
starting and stopping timers, displaying max/min
values and starting/stopping batches. See
Keys & Digital Input Available Settings
a complete list of settings available.
A digital input is available on the meter, and it may
function in a similar fashion as the function keys to
allow remote operation of a single task.
The max & min readings (peak & valley) reached by
the process can be displayed either continuously by
assigning it to a display line in the
Display
menu, or
momentarily by pressing the F1 key (default) or
assign it to any of the other function keys or the digital
input.
Pressing the F1 key cycles through various display
values (e.g. Grand total, max, min); press the Enter
key to lock the display in the current process value
and press the Enter key again to unlock the display.
The relay information menu shows runtime and cycle
count for each relay.
Change the display units within the selected unit class
at any time without the need to re-scale the meter.
Select the desired units via the
UNITS
menu in the
DISPLAY
menu, and the meter automatically converts
the display values to the new unit of measure.
CapTouch Buttons Operation
Button
Description
(Default Settings)
Press to enter or exit
Programming
Mode
, view settings, or exit max/min
readings.
Press to display grand total. Continue
pressing to cycle through max, min,
rate, and total displays.
Press to access the Reset menu.
Press F1 to scroll through the options.
Press F3 to reset the currently
displayed parameter.
Press to acknowledge all manually
resettable relays or open collectors.
Press to lock/unlock the display value
after pressing the F1 key.
CapTouch Button Tips &
Troubleshooting
Use a password to prevent tampering.