PD6928 Ex-Proof & I.S. Loop-Powered Flow Rate/Totalizer
Instruction Manual
18
Compliance Information
Hazardous Area Approvals
CSA
Explosion-proof for use in:
Class I, Division 1, Groups B, C and D
Dust-ignition proof for use in:
Class II/III, Division 1, Groups E, F and G; T6
Flame-proof for use in:
Zone 1, Ex d IIC T6
Ta = -55 to 75°C.
Enclosure: Type 4X & IP66/IP68.
Certificate number: 2325749
ATEX
II 1 G D
Ex ia IIC T4 Ga
Ex ia IIC T200°C Da
Ta = -40 to 75°C
Enclosure: Type 4X & IP66/IP68
Install per Control Drawing DW2636
(contained within
Certificate number: CML 18ATEX2089X
IECEx
II 1 G D
Ex ia IIC T4 Ga
Ex ia IIC T200°C Da
Ta = -40 to 75°C
Enclosure: Type 4X & IP66/IP68
Install per Control Drawing DW2636
(contained within
Certificate number: IECEx CML 18.0050X
ATEX/IECEx Special Conditions for Safe Use
The following conditions relate to safe installation
and/or use of the equipment.
•
The equipment loop/power port shall be connected
to an intrinsically safe barrier with Uo ≥ 5.8V
•
The 4-20 mA input port shall be connected to an
intri
nsically safe barrier with Uo ≥ 5.1V
•
The PD6928-HA-AL enclosure is manufactured from
aluminum. In rare cases, ignition sources due to
impact and friction sparks could occur. This shall be
considered during installation, particularly if the
equipment is installed in a Zone 0 location.
•
All cable entries into the equipment shall be via cable
glands or conduit which provide a minimum degree
of protection of IP54.
•
The equipment may not have 500V isolation
between the circuit and earth. This shall be taken
into account when installing the equipment.
Year of Construction
This information is contained within the serial number
with the first four digits representing the year and
month in the YYMM format.
For European Community:
The PD69XX Series must be installed in accordance
with the ATEX directive 2014/34/EU, the product
certificates CML 18ATEX2089X, IECEx CML
18.0050X and the product manual.
Electromagnetic Compatibility
EMC Emissions
•
CFR 47 FCC Part 15 Subpart B Class A
emissions requirements (USA)
•
ICES-003 Information Technology
emissions requirements (Canada)
•
AS/NZS CISPR 11 Group 1 Class A
ISM emissions requirements
(Australia/New Zealand)
•
EN 55011 Group 1 Class A ISM
emissions requirements (EU)
•
EN 61000-6-4 Emissions requirements
for Heavy Industrial Environments -
Generic
EMC Emissions
and Immunity
EN 61326-1 EMC requirements for
Electrical equipment for measurement,
control, and laboratory use
– industrial use