16 Maintenance and Servicing
■
73
En
gl
is
h
16.8 Error Messages
The moisture analyser shows a description of the fault in the info-line.
16.8.1 Notes on Correcting Faults
The following table shows faults and their possible causes. If you cannot clear the fault using the table,
please contact the Customer Service..
NOTE
If an error occurs without a description of the error in the info-line, the Customer Service must be called.
Fault
Possible Cause
Weight display does
not light
• Moisture analyser not switched on
• Connection to power adaptor is interrupted
• Power supply has failed (interruption to current)
• The power adaptor is defective
“OL” is shown in dis-
play
• The weight range has been exceeded (Observe information on the maximum
weight range)
“UL” is shown in dis-
play
• The weight range is below the range of the moisture analyser (weighin pan or
weighing pan holder missing)
The weight display
fluctuates
continuously
• The draft is too strong at the moisture analyser location
• The moisture analyser support is vibrating or varying
• The scale pan is touching a foreign body
• The refresh rate chosen for is too fast
• The material being weighed is absorbing moisture
• The material being weighed is being blown away or is evaporating
• Strong temperature variations in the material being weighed
Results of weighing
are clearly incorrect
• The moisture analyser was not correctly tared
• The moisture analyser is not correctly levelled
• The adjustment is no longer correct
• There are strong temperature variations
There is no display or
only dashes
• The stability control (Balance functions) is set too sensitively
• The time selected for „FLOATINGDISPLAY“ is unsatisfactory
Configuration menu
cannot be changed
• The password lock is activated in the configuration menu
The display flashes
continuously during
adjustment
• The moisture analyser location is not stable enough (Interrupt adjustment with
«
ON/OFF
» and relocate the moisture analyser in a better position)
• Use of an inprecise calibration weight (only applies to external adjustment)