Application
9.1
Units
.
Assigning the function keys:
9.2
Count
With the aid of the „COUNT” program you can count items of uniform weight (screws, bearings, coins, etc.).
For this, you must first weigh a defined number of items (for example 5 items) and assign the reference num-
ber of pieces to the reference weight so obtained by pressing the corresponding function key.
• SETUP APPLICATION
UNITS
UNIT-2
mg
-----
OFF
Milligram
-----
not active
UNIT-3
GN
-----
OFF
Grain
-----
not active
UNIT-4
ct
-----
OFF
Carat
----
not active
Key
Function
«
g
»
Show measurement in unit 1, e.g. Gram
«
mg
»
Show measurement in unit 2, e.g. Milligramm
«
GN
»
Show measurement in unit 3, e.g. Grain
«
ct
» or «
STO
»
Show measurement in unit 4, e.g. Carat
or Statistics functions if the statistics-program is activ
NOTE
For basic operation, Unit 1 is defined in the Configuration menu (standard unit for all weighing procedures,
if the application program „UNIT“ is not called up, see chapter 8.3.3 "Selecting the weight unit").
Display
Key
Step
+ 8.0700
9
«
F
»
Press until „UNITS“ appears.
Units
+
8070.0
m9
«
B
»
The weight is displayed as milligrams,
9
m9
Gn
ct
By pressing the relevant function key the weight display is switched to the corresponding unit.
The weight is printed in the unit of measurement shown by pressing the
«
»
key.
• SETUP APPLICATION
COUNT
KEY-1
5
KEY-2
10
KEY-3
25
KEY-4
50
Reference-number of pieces = 5
Reference-number of pieces = 10
Reference-number of pieces = 25
Reference-number of pieces = 50
NOTE
Depending on the weight and tolerances of the objects to be counted, you should count a representative
number of items for the regulation of the reference-weight.
Содержание 490 IB Series