8 Operation
14
8.1.2 Operation in the programming mode
The balance can also be operated by remote control. For the corresponding remote control-commands see
chapter 13.2 "Remote control-commands".
8.2
Setting the factory configuration
• Press «
ON/OFF
» to switch the balance on.
• During the start-up process, hold down the «
T
» and «
MENU
» keys until „FACTORY CONFIG.“ appears in
the display, then relase the keys. The balance loads the factory configuration.
8.3
Device configuration
This section explains the structure of the configuration menu and its functions.
The basic adjustment of the balance is defined in the configuration:
• The settings in the sub-paths pre-programmed at the factory are printed here in
bold
.
• For greater clarity, only that part of the menu tree which corresponds to the function is shown with each
description of the function.
• You will find the entire menu tree for the configuration menu in see chapter 15.1 "Configuration menu
tree".
• Explanations of the menu functions are printed in
italics
.
Keys type 1
Keys type 2
Designation
Function in programming mode
«
A
», «
B
»
• Changes within the menu level
«
C
»
«
D
»
• Up/Down movements within the menu
• Changes selected parameters
«
E
»
• Selects parameters
• Stores the changed parameters
«
esc
»
• Interrupts an input
• Leaves the menu
«
ins
»
• Places insert marker (in text entry)
«
clr
»
• Deletes input (in text entry)
«
»
• Inputs a point (in text entry)
Menu
Definable functions
UNIT-1
Unit in which the weighing results are displayed
SET DATA PRINT
Print formats; Type of values to be printed (individual values, continuous printout,
time or load change dependent values, date, time, user, etc.)
SET CALIBRATION
Calibration method
SET
WEIGHING MODE
Stability mode (Quality of the balance location), Auto-Standby mode, zero correc-
tion, tare method (rapid or standard tare)
SET INTERFACE
Baud rate, parity, handshake functions of the peripheral interface
SET
DATE AND TIME
Date and time (standard format or American format p.m. and a.m.)
THEFTCODE
Activating/deactivating and changing the anti-theft code.
KEY TONE
Activation of the keypad sound
LANGUAGE
Display language (E, G, F)
CONTRAST
Screen contrast (only for balances with LCD display)
T
MENU
esc
T
ins
clr
Содержание 490 IB Series