14 Further information
■
59
14 Further information
14.1 Notes on the Weighing mode
14.1.1 Set Weighing mode: Floating Display
The value set for Floating Display defines the period, after which each
new measurement is displayed.
For the definition of this period, the quality of the balance-location is
crucial. The stability control must also be suitably chosen.
Recommended values:
14.1.2 Set Weighing mode: Stability control
The value set for the stability control depends on the quality of the
balance location and must be correctly chosen in order to obtain
optimal, reproducible results. Choose:
• „StAbiL HI“ at an
optimal
balance location,
• „StAbiL MED“ at a
good
balance location or
• „StAbiL LOW“ at a
critical
balance location
• Optimal balance location:
FLd 0.08
• Good balance location:
FLd 0.16
• Critical balance location:
FLd 0.32
NOTE
The value of the Floating Display is a function of the stability con-
trol and the balance location. For balance location, see chapter 3.4
“Choice of a suitable location” and chapter 5.6 “Weighing mode”.