8 Data transfer to peripheral devices
■
49
Tab. 8.4 Remote control-commands
8.4
Examples for the remote control of the
balance
Tab. 8.5 Examples for remote control
Ln
Backlight
n=0
off
n=1
on
n=2
auto
SDTmmddyy
hhmmss
Set Date and Time (English: Month, Day, Year,
Hour, Minutes, Seconds)
Tttt
Tare or set tare to a specific value
Uxnn
Set unit x (1 or 2) of the balance with nn (0=g,
1=mg, 2=kg, ...)
UxS
Switch balance to unit x (1 or 2)
ZERO
Zero balance (provided weight is stable and within
the zero position range)
NOTE
Each remote control-command must terminate with «CR» «LF».
The commands are acknowledged if required.
Input
Description of the function executed
D –––––
Five dashes will be shown
DTEST123
Will show: tESt123
D
The display will be dark
T100
-100.000g (Tare set to 100 g)
T1
-1.00000 g (Tare set to 1 g)
T
Balance will be tared
Command
Function