2.4.3.
Calibrate Channel A (see Figure 6)
Calibrating channel A has one sub-menu.
a. Set Maximum Load.
The maximum load value can be found from the specification of the sensor.
Procedures:
Check current unit settings (Channel A…).
Select
a. Set Maximum Load
and touch
OK
.
Use
UP
or
DOWN
to move cursor.
Touch number key to input the maximum load value in a flashing digit.
Touch
OK
to accept change and cursor will automatically move to
a. Set
Maximum Load.
Or touch
MENU
to return without saving.
2.4.4.
Calibrate Channel B (see Figure 6)
Please refer to
2.4.3. Calibration Channel A
.
2.4.5.
Set Alarms (see Figure 6)
Alarms can work in both Record Mode and Non-Record Mode. However they
can
ONLY
be set in Non-Record Mode (non flashing record icon).
Set Alarms
has ten items.
a. Set Maximum Tension.
b. Set Maximum speed.
c. Set depth alarm1.
d. Set depth alarm2.
e. Set depth alarm3.
f. Set depth alarm4.
g. Set depth alarm5.
h. Set depth alarm6.
i. Set depth alarm7.
j. Set depth alarm8.
Procedures:
Select desired item and touch
OK
.
Use
UP
or
DOWN
to move cursor.
Touch number key to enter desired number in the highlighted digit.
Touch
OK
to accept change and return. Or touch
MENU
to return without
saving.