Pioneer Petrotech Services Inc.
#1, 1431 – 40 Ave NE, Calgary, AB, Canada, T2E 8N6
Tel: (403)282-7669, Fax: (403)282-0509
PPS36 User Manual Page
12
of
30
2.4.
Main Menu
Touch
MENU
button to go into Main Menu.
2.4.1.
Calibrate Depth (see Figure 6)
1) Calibration depth option 1
Calibrating depth option1 has four sub-
menus.
a.
Set Start Depth.
b.
Set Stop Depth.
c.
Calibrate.
d.
Quit Calibration.
Procedures:
Connect the encoder cable to the box
(see Figure 7).
Check current unit settings (depth,
speed …).
Stop running depth system.
Select
a. Set Start Depth
and press
OK
.
Touch
UP
or
DOWN
to move cursor.
Touch number key to input number in a flashing digit.
o
The number can be got by reading an initial mechanical counter or
initial value which the users expect.
o
Check if the desired unit is the same as the DW display unit.
Touch
OK
to accept change and cursor will automatically move to
b. Set
Stop Depth
. Or touch
MENU
to return.
Start running depth system at least 25m or 80 feet or even more if possible
and then stop.
Select
b. Set Stop Depth
and touch
OK
.
Use
UP
or
DOWN
to move cursor.
Touch number key to input number in a flashing digit.
o
The number can be got by reading of mechanical counter or reading
from a measuring ruler plus the initial value.
Touch
OK
to accept change and the cursor will automatically moves to
c.
Calibrate.
Or touch
MENU
to return.
Select
c. Calibrate
and touch
OK
to calibrate, or else touch
MENU
to return.
Select
d. Exit Calibration
and touch
OK
to return.
Figure 6
Figure 7