![PP Systems CIRAS-2 Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/pp-systems/ciras-2/ciras-2_operators-manual_1592667062.webp)
Observing and Recording Leaf Gas Exchange
CIRAS-2 Operator's Manual Version 2.04
- 60 -
Please note:
The CO
2
reference value (Cr) will not be exactly what you set if the CO2 Control Type
selected is (Supply (Ref) Approximate ppm). It should, however, be stable (fluctuating not more than +/- 2
ppm) at a value within approximately 15-30 ppm of the set value. Precise CO
2
control is available to
/- 1 ppm of the set value
by using the „Supply (Ref) Set ppm‟ Control Type.
The system is functioning properly if the above conditions are observed. If not, please refer to the
Troubleshooting Section of the
CIRAS-2 Technical Manual
available in the User section of our website.
Select Recording Options
To begin making measurements, select
Recording
–
Begin
from the CIRAS-2 menu bar.
There are four different methods available for recording data:
1. Keypress
2. Timed
3. Response Curve
4. Batch Response Curves
Содержание CIRAS-2
Страница 2: ...This page is left intentionally blank...
Страница 77: ...Observing and Recording Leaf Gas Exchange CIRAS 2 Operator s Manual Version 2 04 75 Time Plot Data Plot...
Страница 112: ...CIRAS CFM Display CIRAS 2 Operator s Manual Version 2 04 110...
Страница 167: ...User Notes CIRAS 2 Operator s Manual Version 2 04 165 User Notes...