Diagnostics
CIRAS-2 Operator's Manual Version 2.04
- 138 -
Apply Analyser Flow Control
Click after changing analyzer flow setting to establish new flow.
Cuvette Air Supply
Forwarding Flow [A/D] / Cuvette Flow [A/D]
These are the outputs of the respective flowmeters. The forwarding flow should be about 100 greater
than the cuvette flow. Used for testing the cuvette air supply.
Flow (mls/min)
Click on this so that a new value can be entered.
Flow Requested
Flow Rate that CIRAS-2 has been set to.
Flow Measured
Flow rate that is being delivered.
Control Setting
The current setting for the cuvette air supply pump driver. Typically it is a maximum of 113 at 470
mls/minute.
Auto Cuvette
Light Control
Uncorrected PAR is the reading of the PAR sensor. By entering new values for Lamp1 (central lamp in a
PLC6(U)) and Lamp2 (lateral lamps) the intensity can be altered. Used for testing the PAR sensor and
light units.