background image

5

Claw Medical Vacuum Systems

operation

BeFore STarT UP

1.  Make sure all safety warnings, labels and instructions 

have been read and understood before continuing.

2.  Remove any shipping materials, brackets, etc.
3.  Ensure all fuses, circuit breakers, etc., are properly sized.
4.  Verify that all pumps have the proper amount of oil in 

them for operation. The claw type vacuum pumps require 
oil in the gear boxes. See the enclosed manual from the 
pump manufacturer for correct oil type to use.

5.  Confirm electric power source and ground have been 

firmly connected. Make sure the electrical control box 
door is closed and latched.

6.  Make sure inlet filter is properly installed and all piping 

is connected. Open the isolation valves for each vacuum 
pump. Open the valve to the receiver. Close the receiver 
drain valve.

7.  Make sure all selector switches on the control panel are 

OFF.

 

 

Risk of injury.  

  Make sure no one in 

contact with any moving parts during the rotation 
check.

8.  Check that all fuses, circuit breakers, etc., are the proper 

size.

 

If all pumps are running in the  

  wrong direction, change the 

incoming power leads to correct rotation.
 

On some claw vacuum pumps,  

  the fan is powered by a sepa-

rate motor. The fan rotation direction may be different from 
the main shaft rotation direction.

STarT-UP anD oPeraTion

1.  Follow all procedures under “Before Start-Up” before 

attempting operation of the vacuum pump.

2.  Make sure all selector switches are in the OFF position.
3.  Switch on electric source.
4.  Open tank connection valve or valves completely.
5.  Using the selector switches on the control panel, turn 

on each pump – motor in the “Auto” mode until all are 
running.

6.  Check the vacuum level using the system gauge or 

display. The control system will turn off the vacuum 
pump motors when the vacuum level is maintained and 
the minimum run timer conditions are satisfied.

7.  Check for excessive vibration, unusual noises or leaks 

during operation. If problems are detected, shut down 
the system and make corrections or repairs as needed 
before operating the system.

8.  Pumps may be operated in “Hand” mode to override 

the function of the automatic controls. The pumps and 
the system will not be damaged during “hand” mode 
operation. The pumps are equipped with automatic 
vacuum limiting valves if needed to avoid damage.

9.  In normal operation, leave all selector switches in the 

“Auto”position and allow the controls to cycle the pumps 
as needed based on vacuum demand.

 

Switch breaker OFF if vacuum  

  pump will not be used for a 

long period of time.

Sequence of operations

All vacuum pumps will operate on a ten minute minimum 
run timer (to prevent the short cycling of pumps) and 
will alternate Lead, Lag and etc. every 30 minutes, unless 
interrupted by the Lead / Lag Freeze (see example below.)
For example, a vacuum switch closes 25 minutes into the 30 
minute alternation sequence; pump 1 is activated and held 
on for its ten minute minimum run time. The alternation 
sequence will be extended to a total of 35 minutes. During 
the following 30 minutes, if a vacuum switch closes, pump 2 
will be activated and held on for its ten minute run time. This 
operation is referred as the Lead / Lag Freeze. Its purpose is 
to prevent short cycling of pumps in the event that a vacuum 
switch is activated at the end of an alternation period.
The Reserve (or last Lag) Vacuum Switch will bring on the 
reserve or last pump in the system and activate the Reserve 
Pump In Use Alarm. 
The Vacuum Switch’s are factory set to the system 
specifications. It is not recommended to adjust the settings 
in the field. An adjustment to the setting could result in 
inconsistent PLC operation and possible malfunction of the 
system controls. Please consult your service contractor if 
adjustments are required.

Содержание CVPT Series

Страница 1: ...not avoided WILL result in death or serious injury Warning indicates a potentially hazardous situation which if not avoided COULD result in death or serious injury Caution indicates a potentially haz...

Страница 2: ...eiver Tank The receiver tank in a Powerex Medical Vacuum System is plumbed so that it sees the system vacuum It is not pressurized in normal use and should not be connected to any pressure sources The...

Страница 3: ...ctrical supply before servicing the equipment Electrical shock hazard Make sure the system is grounded in accordance with NEC and local requirements All electrical hook ups must be performed by a qual...

Страница 4: ...f the system extends beyond 600 equivalent straight feet use one size larger pipe Approximate system CFM equals the number of pumps running times CFM in table below Data below is for reference only if...

Страница 5: ...he vacuum level is maintained and the minimum run timer conditions are satisfied 7 Check for excessive vibration unusual noises or leaks during operation If problems are detected shut down the system...

Страница 6: ...maintenance schedule relates to the normal operating conditions If the circumstances and load condition are adverse shorten the interval time and perform maintenance accordingly Maintenance LUBRICATIO...

Страница 7: ...eck drive coupling replace if needed 2 Verify that pump shaft turns freely repair or replace as necessary 3 Check vacuum discharge piping to verify air is being exhausted repair or replace as necessar...

Страница 8: ...ns iii that have been materially modified without the prior written approval of Powerex or iv that experience failures resulting from operation either intentional or otherwise in excess of rated capac...

Отзывы: