Claw Medical Vacuum Systems

Troubleshooting Guide

ProBlem

CaUSe

CorreCTiVe aCTion

Lag Alarm

Overload Tripped

1. Reset overload; if problem continues, check motor amp draw
2. Verify overload is set to the correct service factor
3. Verify wire gauge is correct for amp draw

PLC   Failure

1. Check PLC for red error light; replace PLC
2. Check PLC for flashing green light; if 24V DC is present on the last 

input, then reprogram PLC with E-Prom chip

Pump / Motor  Failure

1. Check drive coupling; replace if needed
2. Verify that pump shaft turns freely; repair or replace as necessary
3. Check vacuum discharge piping to verify air is being exhausted; repair 

or replace as necessary

4. Verify voltage to the motor; repair wiring or replace motor as 

necessary

Vacuum Consumption

1. Repair leaks
2. Inspect purge valves are closing; repair or replace as necessary
3. If scroll pump is running backwards, replace check valve

Switch Failure

Check continuity on lead vacuum switch

High Temperature Alarm

High Temperature

Verify ambient temperature is below 104 ° F

Lost Connection

Inspect high temperature probe and connections; replace if necessary

notes

Содержание CVPT Series

Страница 1: ...not avoided WILL result in death or serious injury Warning indicates a potentially hazardous situation which if not avoided COULD result in death or serious injury Caution indicates a potentially haz...

Страница 2: ...eiver Tank The receiver tank in a Powerex Medical Vacuum System is plumbed so that it sees the system vacuum It is not pressurized in normal use and should not be connected to any pressure sources The...

Страница 3: ...ctrical supply before servicing the equipment Electrical shock hazard Make sure the system is grounded in accordance with NEC and local requirements All electrical hook ups must be performed by a qual...

Страница 4: ...f the system extends beyond 600 equivalent straight feet use one size larger pipe Approximate system CFM equals the number of pumps running times CFM in table below Data below is for reference only if...

Страница 5: ...he vacuum level is maintained and the minimum run timer conditions are satisfied 7 Check for excessive vibration unusual noises or leaks during operation If problems are detected shut down the system...

Страница 6: ...maintenance schedule relates to the normal operating conditions If the circumstances and load condition are adverse shorten the interval time and perform maintenance accordingly Maintenance LUBRICATIO...

Страница 7: ...eck drive coupling replace if needed 2 Verify that pump shaft turns freely repair or replace as necessary 3 Check vacuum discharge piping to verify air is being exhausted repair or replace as necessar...

Страница 8: ...ns iii that have been materially modified without the prior written approval of Powerex or iv that experience failures resulting from operation either intentional or otherwise in excess of rated capac...

Отзывы: