Powered by Safety
®
25
Installation
01.4IB.48040C
Insert the fibers individually into the
optical connectors until they come
to a hard stop. The stop will be very
noticeable and does not require
excessive force. Secure the Fiber
Retention Nut with moderate force by
screwing it in clockwise as shown in
. Do not over tighten the nut,
as it is plastic and can be broken if too
much force is applied.
Figure 31 Fiber Installation
d. Ensure that Optical Signal is adequate
once all the fibers are installed and
connected, they must be checked to
ensure that the optical signal is strong
enough for proper operation of the
device.
i. Apply power to the device and wait
several minutes for all probes to
stabilize.
ii. For each probe, look at the color
of the BriteSpot™ Status LED and
follow the actions recommended in
Table G BriteSpot™ Status LED Colors
.
Table G BriteSpot™ Status LED Colors
Status LED Color
Status
Action
Green
OK
No Action
Red
Fail
Trim several mm off
of fiber end,
re-insert
iii. Complete the procedure by
checking the fiber power for the
remaining probes. Read the fiber
power screen (only BSG3.9S) or read
fiber power registers (BSG3.9 and
BSG3.18). Fiber power shall be 7,8,9
or 10 (strongest signal)
iv. If any of the probes continue to
fail, see
for
remedial actions.
Содержание BriteSpot BSG3
Страница 1: ...Instruction Bulletin 01 4IB 48040C BriteSpot BSG3 Powered by Safety ...
Страница 4: ...Powered by Safety BriteSpot BSG3 01 4IB 48040C This page is intentionally left blank ...
Страница 10: ...Powered by Safety BriteSpot BSG3 01 4IB 48040C This page is intentionally left blank ...
Страница 37: ...Powered by Safety 27 Usage 01 4IB 48040C Figure 34 LCD Menu and Navigation ...