Fault Finding
WARNING: THIS EQUIPMENT SHOULD ONLY BE DISMANTLED BY PROPERLY TRAINED
PERSONNEL. REMOVING THE MAIN COVER EXPOSES POTENTIALLY LETHAL MAINS VOLTAGES.
Common Faults
• Nothing happens when the unit is
switched on
• The unit turns on but the gas does not
heat
• No gas emits from the needles
• Certain wells do not dry down
• The instrument does not move to it’s set
position when running a method
• The instrument continues to emit gas
after a method has finished.
Solution
o
Check power to the unit. Check the condition of
the fuses.
o
Check that the inlet gas pressure is greater than
4 Bar (60 psi). Turn the unit off and on again to
reset heater.
o
Check that the gas inlet supply is sufficient.
Check integrity of head gasket. Check for gas
leaks.
o
Check visually that the needles are not blocked.
Check visually that the needles are not bent or
damaged. Due to the nature of the heated gas,
the pins towards the centre of the head plate will
generally be slightly warmer.
o
Check that the load and eject positions are set
correctly in the Setup Position menu.
o
Turn the unit on and off to reset the gas
regulation.
Special Note
For other failures it is recommended that you contact your supplier. Do not dismantle the unit, as
this will invalidate the warranty.