Page 20
V 1.3 – February 2016
U s e r I n s t r u c t i o n M a n u a l
Fitting / Changing Needle Head
Before removing / changing the needle head, disconnect the unit from any electrical
supplies and gas. Also ensure Ultravap Mistral has cooled down sufficiently if used.
• Remove the safety guard.
• To remove the needle head, remove the two fixing screws, using the Allen key provided.
Turning the unit on it’s side may make this process easier (ensure it is placed on a suitable
surface and a cloth is used if necessary for protection).
• Carefully remove the head, avoid bending or damaging needles.
• Inspect gasket for tears and replace if needed.
• When replacing the needle head, great care must be taken to ensure that the gasket is
present and fitted correctly. Also, ensure that the needle head locates on the dowels and the
opposing corner screws are used so that the needle head is fixed evenly. For ease of fitting;
the needle head will fit either way round.
Fixing screw
Fixing screw