Setting Evaporation Parameters
In addition to the standard parameters in the form
P%=###<cr> and P%<cr>
The following parameters are possible in the form
P%%=###<cr> and P%%<cr>
Parameter
Description
Units
Limits
01
*Gas Temperature 1
˚C
1 to 60
02
*Gas Temperature 2
˚C
1 to 60
03
*Gas Temperature 3
˚C
1 to 60
04
*Gas Temperature 4
˚C
1 to 60
05
*Gas Temperature 5
˚C
1 to 60
06
Distance 1
0.25mm
1 to 220
07
Distance 2
0.25mm
1 to 220
08
Distance 3
0.25mm
1 to 220
09
Distance 4
0.25mm
1 to 220
10
Distance 5
0.25mm
1 to 220
11
Distance 6
0.25mm
1 to 220
12
Time 1
Minutes
1 to 60
13
Time 2
Minutes
1 to 60
14
Time 3
Minutes
1 to 60
15
Time 4
Minutes
1 to 60
16
Time 5
Minutes
1 to 60
17
**Gas flow 1
5L/min
1 to 18
18
**Gas flow 2
5L/min
1 to 18
19
**Gas flow 3
5L/min
1 to 18
20
**Gas flow 4
5L/min
1 to 18
21
**Gas flow 5
5L/min
1 to 18
Note *For Gas Temperature the Mistral has no direct cooling facility, The Mistral`s lowest achievable
temperature is to ambient
Note ** For Gas flow the Low Flow Error may appear if the Flow rate is set to < than 20 L/min also the
temperature will not increase if the flow is set to < than 30 L/min.
Remote Communications Protocol