IMPORTANT: 

READ MANUAL COMPLETELY Be f o r e   O p e r a t i n g   T h i s   D e v i c e  

Basic delivery technique is described. Also, this manual contains instructions on periodically required checks to be 
performed by the user.  These checks are necessary to insure the proper performance of this device and its safety 
features.  Retain this manual for future reference. 

 

 

WARNINGS AND PRECAUTIONS 

 

These warnings and precautions are to help you to 
understand how to safely operate the PORTER Digital 
MDM Flowmeter.  A WARNING alerts you to a 
possible hazard to people.  A CAUTION alerts you to 
the possibility of equipment damage. 

WARNING:

  Do not use this device for the 

administration of general anesthesia or as a part of, or 
in conjunction with, a general anesthesia 
administration system. 

WARNING:

 Workers are exposed to N

2

O during 

administration of N

2

O/O

conscious sedation 

analgesia.  NIOSH has recommended that exposures 
should be minimized.  Contact NIOSH (1-800-35-
NIOSH) to receive NIOSH Publications on 

Control of 

Nitrous Oxide in Dental Operatories.  

Exposure can be 

minimized by effective controls.  National Institute for 
Occupational Safety and Health (NIOSH) publications 
state that controls, including System Maintenance, 
Ventilation and Work Practices can effectively reduce 
N

2

O concentrations in patient procedures.  Your 

flowmeter accessory Porter scavenger system is an 
important part of the system of controls. 

 

WARNING:

  Porter Instrument equipment utilizes the 

cross+protection 

system. The flexible hose and 

connectors that connect to the flowmeter are diameter 
indexed; 3/8” O.D. for Nitrous Oxide and ½” for 
Oxygen. The 

cross+protection 

system is designed to 

prevent misconnection of Oxygen and Nitrous Oxide 
piping.  

DO NOT ATTEMPT TO CHANGE THE 

DIAMETERS OR CONNECTORS OF THE 
FLOWMETER!  

Tampering with the 

cross+protection

 

system constitutes acceptance of liability by the 
installer. For your own protection, as well as that of the 
Doctor and the patients, use 3/8” O.D. tubing for all 
Nitrous Oxide lines and ½” O.D. tubing for all Oxygen 
lines.  

To assure safe operation and conformation to local fire 
codes, all Porter Instrument flowmeter systems are 
designed to be used with sedation delivery systems 
mounted inside walls and they meet or exceed the 

guidelines established by the National Fire Protection 
Association for Nonflammable Medical Gas Systems, 
NFPA 99. Copies of NFPA 99 or portions thereof may 
be obtained by writing to:  National Fire Protection 
Association, Batterymarch Park, Quincy, MA 02269-
9904; or call 1-800-344-3555 

WARNING:

  New or modified installations - 

properly 

connected gas pipelines are absolutely essential 
to patient safety.

  The authorized distributor or 

contractor should provide written documentation that 
all gas pipelines are connected properly and that the 
system has been pressure tested prior to use.  While 
this is a good business practice, it is important that the 
user verify by their own test, independent of the 
authorized distributor or contractor, that all gas 
pipelines are connected correctly prior to using the 
system.  

The ultimate responsibility of assuring 

that lines are not crossed rests with the user. 

WARNING:  

During any power outage, remember to 

turn OFF the flowmeter and manually turn OFF the 
tank valves.   With centralized, electrically powered 
gas systems, if gas was flowing when the power went 
out and the flowmeter is left ON, gas will be flowing 
when the power is restored. 

 

CAUTION:

  This device is for use by or on the order of 

a physician, dentist or licensed healthcare 
professional. 
 

CAUTION:

 Always use clean, dry medical grade 

gases.  Introduction of moisture or other contaminants 
into this device may result in defective operation. 
 

CAUTION: 

Do not attempt to repair, alter or calibrate 

this device.  Unauthorized repair, alteration or misuse 
of this device is likely to adversely affect the 
performance and will void the warranty. 

 

CAUTION: 

Never oil or grease any part of this 

system (minimize fire or explosion potential).

 

 

Содержание DIGITAL MDM

Страница 1: ...ration is The design manufacture distribution and servicing of Nitrous Oxide Oxygen Sedation Flowmeters Gas Scavenging Systems Gas Distribution Systems and Office Communication Systems for use by a ph...

Страница 2: ...own protection as well as that of the Doctor and the patients use 3 8 O D tubing for all Nitrous Oxide lines and O D tubing for all Oxygen lines To assure safe operation and conformation to local fire...

Страница 3: ...A known Protective Earth ground connection must be made via one of the following methods 1 The third wire on the power cord when plugged into an appropriate outlet that is properly installed and grou...

Страница 4: ...ropriate regulatory authority in your country state or province Training is recommended to provide a practical hands on capability and an understanding of the behavioral aspects of Nitrous Oxide Sedat...

Страница 5: ...or pipeline gas systems and delivers the electronically controlled mixture to the bag tee outlet Built In Safety A series of alarms see Safety Alarm System page 6 alert the user to any problem conditi...

Страница 6: ...n restored either by disconnecting reconnecting the power cord or by local power failure the unit will display OFF The operator must press the ON OFF key to restart the unit The unit will then display...

Страница 7: ...leasing the key returns the unit to the flow rate and mixture ratio which were in effect prior to the flush Alarm Test Silence Key The alarm test silence key serves two functions When pressed all the...

Страница 8: ...8 PORTER DIGITAL MDM SYSTEM ILLUSTRATION...

Страница 9: ...alve 9A B 2461 001 N2O Check Valve 10 8005 O2 Hose Diss Diss 5 Ft 10A 8505 N2O Hose Diss Diss 5 Ft 10B 80012 O2 Hose Diss Diss 12 Ft 10C 85012 N2O Hose Diss Diss 12 Ft 11 PA 450 1 002 O2 Hose 12 Ft 11...

Страница 10: ...ews again Bracket B 5485 Figure 3 Control Unit mount directly to wall 3 Use bracket shown in Figure 4 for under cabinet mounting Turn same bracket upside down not shown for table top mount Kit B 5555...

Страница 11: ...NLET fitting on the back of the mixer head see Figure 6 Hand tighten and then finish tightening with a 7 8 inch open end wrench approximately 1 8 turn Do not over tighten 2 Connect the DISS female hex...

Страница 12: ...12 Figure 6 Connection to the back of unit Figure 7 Installation of Power Cord Figure 8 Connections Schematic some components may not be part of your installation Power Cord N2O DISS O2 DISS...

Страница 13: ...servoir bag to the bottom downspout of the bag tee Attach the vacuum hoses to the vacuum control source AVS or In line Vacuum Control Block Refer to FM 809 for User Instructions Assembly Installation...

Страница 14: ...g the cut ends of the tubing to both ends of the vacuum control block Adjust the vacuum flow using the control knob Vacuum flow is most effective when the ball float is set within the green bar area R...

Страница 15: ...ve See Fig 10 below is not functioning properly and should be replaced Figure 10 Emergency Air Valve Check Monthly Place the On Standby switch in the STANDBY Off position Disconnect the corrugated rub...

Страница 16: ...1 to 4 years requires caution and specific protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical p...

Страница 17: ...n Failure Test Set the LPM FLOW display to 9 9 LPM Set the OXYGEN display to 50 Turn off the oxygen gas supply to the unit When the gas runs out check for each of the following the OXYGEN FAILURE alar...

Страница 18: ...Handbook of Nitrous Oxide and Oxygen Sedation written by Clark and Brunswick and published by Mosby www mosby com This text is a concise and contemporary guide for nitrous oxide oxygen administration...

Страница 19: ...nimum Nitrous Oxide Concentration 00 Maximum Nitrous Oxide Concentration 70 Minimum Oxygen Concentration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture...

Отзывы: