18 

 

Indicated Flow Delivery Test:

 

  

Set the % OXYGEN display to 100%. Press the 

[

]

 

flow key and hold, as shown in Figure 1. The 

flow rate should decrease to 1.0 LPM. An audible 
alarm sounds at the lower limit while the key is 

held. Press the [

]

 

flow key and hold. The flow 

rate should increase to 9.9 LPM. An audible alarm 
will sound at the upper limit as long as the key is 
held. 
 

Indicated Percentage (%) Delivery Test:

  

Set the LPM FLOW display to 9.9 LPM. Press and 

hold the [

] % OXYGEN key. The % OXYGEN 

display should decrease to 30%. The audible 
alarm sounds at the lower limit while the key is 

held. Press and hold the [

] % OXYGEN key, as 

shown in Figure 1. The % OXYGEN display 
should increase to 100%. An audible alarm 
sounds at the upper limit while the key is held. 
 

Basic Delivery Technique: 

Practice titration.  Titration is a method of 
administering a substance by adding definitive 
amounts of a drug until an endpoint is reached.  
For nitrous oxide / oxygen (N

2

O) / O

2

) sedation, 

N

2

O is given in incremental doses until a patient 

has reached a comfortable relaxed state of 
sedation.  The ability to titrate N

2

O is a significant 

advantage because it limits the amount of drug to 
that which is required by the patient.  If titration is 
done properly, the patient does not receive any 
more of the drug than is necessary.  The amount 
of N

2

O required by a patient on any given day or 

time varies. 
  
For information on titration, a most valuable 
resource for the practitioner is the Handbook of 
Nitrous Oxide and Oxygen Sedation, written by 
Clark and Brunswick and published by Mosby 
(www.mosby.com).  This text is a concise and 
contemporary guide for nitrous oxide / oxygen 
administration. 

 
Maintenance 

The DIGITAL MDM mixer is a precision analgesia 
mixing unit. As with any medical instrument, it 
requires periodic inspection and maintenance. 
The following routine maintenance procedures 
should be performed to ensure system integrity 
and trouble-free service. 

-Inspect mixer, hoses, and connections daily for 
damage, wear, and leaks. 

-Perform functional tests prior to each use. These 
tests are described in Performance Checks,  

-While cleaning the outside of the unit, observe 
caution so that liquid does not get inside. Porter 
recommends using a damp (wrung toward dry) 
cloth moistened with a mild soap or a plastic-safe 
cleaner. Do not spray directly on the case. A hard 
surface disinfectant may be applied in the same 
manner. 

CAUTION: DO NOT submerge the 

unit—use a damp cloth

 

 

WARNING: Before cleaning the unit, 

disconnect the DIGITAL MDM mixer’s power 
cord from the electrical power supply. 
 
 

Fuse Replacement in Domestic, 115-120V 
and International, 220-240V 

1.  To replace a blown fuse, turn the unit OFF, 

disconnect power to the unit by unplugging 
the power cord from the wall, and then 
remove the power cord from the input 
connector on the rear of the unit. 

2.  Open the fuse compartment, as shown in 

Figure 11.  

3.  Remove each blown fuse from its clip and 

replace it with a 5mm x 20mm cartridge fuse, 
rated 2.5 amp, 250 volt, slo-blo (time delay), 
part number 63882537. Two fuses are 
required for the unit to operate. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Figure 11  Fuse Replacement

 

 
 
 
 
 

Содержание DIGITAL MDM

Страница 1: ...ration is The design manufacture distribution and servicing of Nitrous Oxide Oxygen Sedation Flowmeters Gas Scavenging Systems Gas Distribution Systems and Office Communication Systems for use by a ph...

Страница 2: ...own protection as well as that of the Doctor and the patients use 3 8 O D tubing for all Nitrous Oxide lines and O D tubing for all Oxygen lines To assure safe operation and conformation to local fire...

Страница 3: ...A known Protective Earth ground connection must be made via one of the following methods 1 The third wire on the power cord when plugged into an appropriate outlet that is properly installed and grou...

Страница 4: ...ropriate regulatory authority in your country state or province Training is recommended to provide a practical hands on capability and an understanding of the behavioral aspects of Nitrous Oxide Sedat...

Страница 5: ...or pipeline gas systems and delivers the electronically controlled mixture to the bag tee outlet Built In Safety A series of alarms see Safety Alarm System page 6 alert the user to any problem conditi...

Страница 6: ...n restored either by disconnecting reconnecting the power cord or by local power failure the unit will display OFF The operator must press the ON OFF key to restart the unit The unit will then display...

Страница 7: ...leasing the key returns the unit to the flow rate and mixture ratio which were in effect prior to the flush Alarm Test Silence Key The alarm test silence key serves two functions When pressed all the...

Страница 8: ...8 PORTER DIGITAL MDM SYSTEM ILLUSTRATION...

Страница 9: ...alve 9A B 2461 001 N2O Check Valve 10 8005 O2 Hose Diss Diss 5 Ft 10A 8505 N2O Hose Diss Diss 5 Ft 10B 80012 O2 Hose Diss Diss 12 Ft 10C 85012 N2O Hose Diss Diss 12 Ft 11 PA 450 1 002 O2 Hose 12 Ft 11...

Страница 10: ...ews again Bracket B 5485 Figure 3 Control Unit mount directly to wall 3 Use bracket shown in Figure 4 for under cabinet mounting Turn same bracket upside down not shown for table top mount Kit B 5555...

Страница 11: ...NLET fitting on the back of the mixer head see Figure 6 Hand tighten and then finish tightening with a 7 8 inch open end wrench approximately 1 8 turn Do not over tighten 2 Connect the DISS female hex...

Страница 12: ...12 Figure 6 Connection to the back of unit Figure 7 Installation of Power Cord Figure 8 Connections Schematic some components may not be part of your installation Power Cord N2O DISS O2 DISS...

Страница 13: ...servoir bag to the bottom downspout of the bag tee Attach the vacuum hoses to the vacuum control source AVS or In line Vacuum Control Block Refer to FM 809 for User Instructions Assembly Installation...

Страница 14: ...g the cut ends of the tubing to both ends of the vacuum control block Adjust the vacuum flow using the control knob Vacuum flow is most effective when the ball float is set within the green bar area R...

Страница 15: ...ve See Fig 10 below is not functioning properly and should be replaced Figure 10 Emergency Air Valve Check Monthly Place the On Standby switch in the STANDBY Off position Disconnect the corrugated rub...

Страница 16: ...1 to 4 years requires caution and specific protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical p...

Страница 17: ...n Failure Test Set the LPM FLOW display to 9 9 LPM Set the OXYGEN display to 50 Turn off the oxygen gas supply to the unit When the gas runs out check for each of the following the OXYGEN FAILURE alar...

Страница 18: ...Handbook of Nitrous Oxide and Oxygen Sedation written by Clark and Brunswick and published by Mosby www mosby com This text is a concise and contemporary guide for nitrous oxide oxygen administration...

Страница 19: ...nimum Nitrous Oxide Concentration 00 Maximum Nitrous Oxide Concentration 70 Minimum Oxygen Concentration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture...

Отзывы: