background image

 

 

Flow Display Control:

 The key pad DISPLAY O

2

 or 

DISPLAY N

2

O select the gas flow rate displayed on the 

flow display. Press and hold the DISPLAY N

2

O key to 

display the nitrous oxide component of the total flow. 
Press and hold the DISPLAY O

2

 key to display the 

oxygen component of the total flow. If neither key is 
held, the LPM FLOW display indicates the total gas 
flow. 
 

Oxygen Flush Key:

 Pushing the O

2

 FLUSH key 

delivers 100% oxygen directly to the mixed gas outlet 
at the rear of the unit. This outlet is connected to the 
patient through the breathing tube connector (see 
Figure 2). While the key is pushed, 100% oxygen is 
delivered continuously at a minimum flow rate of 20 
LPM. Releasing the key returns the unit to the flow rate 
and mixture ratio which were in effect prior to the flush. 
 

Alarm Test/Silence Key:

 The alarm test / silence 

key serves two functions. When pressed, all the 
display, indicators and the audible alarm turns ON. 
This key also silences the audible alarm when a 
NITROUS FAILURE condition exists. The audible 
alarm will be silent for about five minutes and then 
resume, unless the failure condition has been 
corrected. 
 

 

Machine Diagnostics 

The DIGITAL MDM mixer performs self checks at start 
up and during operation. If any errors are detected, the 
unit displays EE on the % OXYGEN display and an 
error code on the LPM FLOW display. If an error code 
appears, remove the unit from patient service and call 
the Porter Technical Service Department at 1-888-723-
4001.  

To aid in the diagnostics of the unit, make note of the 
final error code displayed. 

 
 

Содержание DIGITAL MDM

Страница 1: ...ration is The design manufacture distribution and servicing of Nitrous Oxide Oxygen Sedation Flowmeters Gas Scavenging Systems Gas Distribution Systems and Office Communication Systems for use by a ph...

Страница 2: ...own protection as well as that of the Doctor and the patients use 3 8 O D tubing for all Nitrous Oxide lines and O D tubing for all Oxygen lines To assure safe operation and conformation to local fire...

Страница 3: ...A known Protective Earth ground connection must be made via one of the following methods 1 The third wire on the power cord when plugged into an appropriate outlet that is properly installed and grou...

Страница 4: ...ropriate regulatory authority in your country state or province Training is recommended to provide a practical hands on capability and an understanding of the behavioral aspects of Nitrous Oxide Sedat...

Страница 5: ...or pipeline gas systems and delivers the electronically controlled mixture to the bag tee outlet Built In Safety A series of alarms see Safety Alarm System page 6 alert the user to any problem conditi...

Страница 6: ...n restored either by disconnecting reconnecting the power cord or by local power failure the unit will display OFF The operator must press the ON OFF key to restart the unit The unit will then display...

Страница 7: ...leasing the key returns the unit to the flow rate and mixture ratio which were in effect prior to the flush Alarm Test Silence Key The alarm test silence key serves two functions When pressed all the...

Страница 8: ...8 PORTER DIGITAL MDM SYSTEM ILLUSTRATION...

Страница 9: ...alve 9A B 2461 001 N2O Check Valve 10 8005 O2 Hose Diss Diss 5 Ft 10A 8505 N2O Hose Diss Diss 5 Ft 10B 80012 O2 Hose Diss Diss 12 Ft 10C 85012 N2O Hose Diss Diss 12 Ft 11 PA 450 1 002 O2 Hose 12 Ft 11...

Страница 10: ...ews again Bracket B 5485 Figure 3 Control Unit mount directly to wall 3 Use bracket shown in Figure 4 for under cabinet mounting Turn same bracket upside down not shown for table top mount Kit B 5555...

Страница 11: ...NLET fitting on the back of the mixer head see Figure 6 Hand tighten and then finish tightening with a 7 8 inch open end wrench approximately 1 8 turn Do not over tighten 2 Connect the DISS female hex...

Страница 12: ...12 Figure 6 Connection to the back of unit Figure 7 Installation of Power Cord Figure 8 Connections Schematic some components may not be part of your installation Power Cord N2O DISS O2 DISS...

Страница 13: ...servoir bag to the bottom downspout of the bag tee Attach the vacuum hoses to the vacuum control source AVS or In line Vacuum Control Block Refer to FM 809 for User Instructions Assembly Installation...

Страница 14: ...g the cut ends of the tubing to both ends of the vacuum control block Adjust the vacuum flow using the control knob Vacuum flow is most effective when the ball float is set within the green bar area R...

Страница 15: ...ve See Fig 10 below is not functioning properly and should be replaced Figure 10 Emergency Air Valve Check Monthly Place the On Standby switch in the STANDBY Off position Disconnect the corrugated rub...

Страница 16: ...1 to 4 years requires caution and specific protocols developed by the medical professional upper age limit at determination of medical professional Weight limitations at the determination of medical p...

Страница 17: ...n Failure Test Set the LPM FLOW display to 9 9 LPM Set the OXYGEN display to 50 Turn off the oxygen gas supply to the unit When the gas runs out check for each of the following the OXYGEN FAILURE alar...

Страница 18: ...Handbook of Nitrous Oxide and Oxygen Sedation written by Clark and Brunswick and published by Mosby www mosby com This text is a concise and contemporary guide for nitrous oxide oxygen administration...

Страница 19: ...nimum Nitrous Oxide Concentration 00 Maximum Nitrous Oxide Concentration 70 Minimum Oxygen Concentration 30 Maximum Oxygen Concentration 100 Flow Range 100 Oxygen 1 0 to 9 9 LPM Flow Range Any Mixture...