9.4.2 Configuring relays in mode 2:
In the above example, the relays are configured as indicated below:
50
40
R1
R2
-140
-150
Alarm
threshold
R2
Hysteresis
Hysteresis
Alarm
threshold
R1
Inactive
Active
Inactive
Inactive
Inactive
Active
Active
Active
Use the
and
keys to
increase the values.
Use the
and
keys to
navigate the selections.
Configure the two relays and press ENTER.
CONFIG
MODE 2: ADJUSTMENT OF THE
THRESHOLDS AND HYSTERESI
Rel 1 connected to redox measurement:
Alarm threshold: -150mV Hyst: 10mV
Below
the threshold, R1 is on
Rel 2 connected to redox measurement:
Alarm threshold: 50mV Hyst: 10mV
Above the threshold, R2 is on
CONFIG
WHAT DO YOU WANT TO DO?
Averaging sensor response
Graphic screen
4-20mA outputs adjustment
Relay adjustment
Measure units
Language
Factory default value reminder
Esc: Cancel the procedure
Enter: Validate the choice