User manual
PONSEL
Ref.: NOTICE_ACTEON_2020_v003
V : 003
37/57
9.1 Configuring sensor response averaging
This configuration makes the sensor response more stable and responsive.
By default, the averaging procedure involves 10 measurements. This means that the displayed value is the
average measurement over the last 10 seconds:
Comment :
Increase the averaging if your measurement is unstable.
Decrease the averaging if the measurement process is too slow.
Use the
and
keys to select
the configuration required.
Select the
Averaging sensor response
menu and press
ENTER.
Use the
and
keys to increase
and decrease the averaging.
Select the averaging desired and press ENTER.
CONFIG
CONFIG
WHAT DO YOU WANT TO DO?
Averaging sensor response
Graphic screen
4-20mA outputs adjustment
Relay adjustment
Measure units
Language
Factory default value reminder
AVERAGING SENSOR RESPONSE
1 = Instantaneous answer
100 = Answer very strongly averaged
CONFIG
WHAT DO YOU WANT TO DO?
Averaging sensor response
Graphic screen
4-20mA outputs adjustment
Relay adjustment
Measure units
Language
Factory default value reminder
Average on:
10
measures.
Esc: Cancel the procedure
Enter: Validate the averaging