6.1.2 Redox sensor slope adjustment
If you observe a small measurement error, you may only adjust the slope of your sensor by carrying out the
following procedure:
CALIBRATION
SENSOR CALIBRATION
ORP SENSOR
Gain adjust
Complete calibration
Theoretical calibration
TEMPERATURE SENSOR
Gain adjust
Complete calibration
Theoretical calibration
ENTER: Validate the choice
Use the
and
keys to select
the type of calibration and the
probe to be calibrated.
Select the
Gain adjust
menu and press
ENTER
.
Use the
and
keys to
adjust the buffer value (0mV to
999mV)
Press the ENTER key to confirm gain adjust.
AWAIT MEASUREMENT STABILISATION
Immerse the Redox sensor in the 240mV or 470mV
ORP standard solution.
CALIBRATIO
N
ORP SENSOR
Immerse the sensor in
an ORP standard solution
Standard:
Measure:
mV
mV
ESC: Leave the current slope
ENTER: Validate the new slope
240
234
WAIT: Nonstable measurement