6.1.1 Two point Redox (ORP) sensor calibration (complete calibration):
Use the following procedure to calibrate your Redox sensor.
You will need the 240mV or 470mV Redox buffer solution for this procedure.
CALIBRATION
SENSORS CALIBRATION
ORP SENSOR
Gain adjust
Complete calibration
Theoretical calibration
TEMPERATURE SENSOR
Gain adjust
Complete calibration
Theoretical calibration
ENTER: Validate the choice
Use the
and
keys to
select the type of calibration
and the probe to be calibrated.
Select the
Complete calibration
menu and press
ENTER.
Use the
and
keys to move the
cursor in the name section.
Use the
and
keys to change the
letters. (The scrolling order of the
letters is A…Z,0..9,?,>,space )
Enter your name or reference then press ENTER
CALIBRATION
ORP SENSOR
ESC: Cancel the procedure
ENTER: Start calibration
Enter your name:
D
urant