1-3
1 Safety Information
1.2.5
Cleaning
WARNING!
Electric shock caused by penetrating moisture!
If the devices are live, penetrating moisture can cause an electric shock.
» Always disconnect the corresponding mains plug before cleaning devices.
ATTENTION!
Damage to the housing surfaces caused by using unsuitable cleaning
agents!
» Do not use any aggressive solvents such as acetone or ethanol.
1.2.6
Installing Other Components
ATTENTION!
Impairment of the system caused by installing third-party hardware or
software components!
Off-system hardware or software components could damage the system or
impair the function of the software. This also applies for the change of
hardware drivers which were not provided by Polytec and the change of
configurations of the driver options.
» Do not update installed drivers, especially when updating via Windows®
update.
» Do not run optional updates.
» If you want to install components which have not been distributed by
Polytec, please contact Polytec first.
1.2.7
Disposal
Device
Unless otherwise specified, instruments and their components which are
defective or no longer required must be properly disposed in accordance with
the applicable legal provisions.
Component
Cleaning Agents
Scanner mirrors (dust)
Bellows
Housing surfaces
Mild detergents or disinfectant solutions