SODIMAT - Model 9073
6-1
REMARK
Any result (calibration or measurement) is always brought back to the reference temperature
(25
EE
C, 77
EE
F). If the sample temperature is different from the reference temperature, it is
required to execute a manual or automatic temperature compensation.
NOTE
See chapter 5 for Programming the analyser.
Chapter 6 : Calibrating the instrument
6.1 Calibration of the flowmeter
Calibrate the flowmeter before any other operation. It is also required to adjust the flow
measurement which has been already factory-preadjusted but which should be adjusted on
site.
Procede as follows :
- Measure the output flowrate (weighing of water to drain during 10 minutes, for instance)
- Enter the programming mode and scroll up to command C101 : flowrate calibration, enter
the value. (See chapter 5 : programming for further information).
6.2 Calibration of the temperature sensor
The temperature sensor is located 1 cm under the measurement electrode. It is factory-
preadjusted but needs to be calibrated in the sample on site. This calibration must be
realised before the sodium measurement calibration.
To execute this calibration remove carefully the measurement electrode and replace it by a
precision thermometer in order to obtain a reference value.
6.2.1 Temperature automatic compensation
The sensor measures continuously the sample temperature. The concentration values are
automatically calculated according to the reference temperature (25
E
C) by a pre-
programmed law in the transmitter.
The procedure is as follows (you can also check the commands in Chapter 5) :
Содержание SODIMAT 9073
Страница 1: ...SODIMAT Model 9073 Instruction manual 03 2001 621 190 073 Rev D ...
Страница 2: ......
Страница 12: ......
Страница 20: ......
Страница 27: ...SODIMAT Model 9073 3 7 3 3 PREAMPLIFIER Figure 3 4 Preamplifier connections ...
Страница 34: ......
Страница 36: ...SODIMAT Model 9073 5 2 5 1 General configuration and calibrations ...
Страница 38: ...SODIMAT Model 9073 5 4 5 2 Alarm configuration ...
Страница 40: ...SODIMAT Model 9073 5 6 5 3 Analogue output configuration ...
Страница 42: ...3 2 1 0 0 1 1 0 2 3 4 5 SODIMAT Model 9073 5 8 5 4 Serial interface configuration ...
Страница 56: ......
Страница 58: ......
Страница 65: ...SODIMAT Model 9073 10 4 Figure 10 2 Mixed bed for AUTOCAL version 1 Mixed bed cartridge 2 Control valve ...
Страница 69: ...SODIMAT Model 9073 10 8 Figure 10 4 Changing the pump tubing ...
Страница 74: ......
Страница 80: ......