electrolyte
SODIMAT - Model 9073
3-9
î
Put the electrode into its compartment ;
Do not block the diffusion ring in the cell.
(see Figure
î
).
- Screw the reference electrode in its place.
The electrode tightness should ensure a
slight squashing of the washer (0.5 mm
maximum).
L
Never force it
!
Figure
î
î
ï
Remove the connector cap and connect
the cable (thin wire with AS7 connector) by
screwing till tightness on the washer.
- Mount the electrolyte feeding pipes (see
Figure
ï
ï
).
Figure
ï
ï
- sample height in the measuring cell
ð
Fill in half the container with electrolyte
(See Figure
ð
).
-Open very slightly the lower feeding pipe
and squeeze it to let the electrolyte current
down to the electrode (use absorbent paper
to sponge the electrolyte).
-Open very slightly the upper feeding pipe
and squeeze it to let the electrolyte current
down to the electrode (use absorbent paper
to sponge the electrolyte).
Figure
ð
ð
- If necessary add electrolyte (3/4 of the
container) and put its cap back..
3.8 Floading and starting the analyser
Before flooding, check the adjusting valve is
closed (clockwise) (Figure 2-1 or 2-3).
L
Flooding should be done before
grounding
Flood the instrument, adjust the flowrate to
4 l/h and check the following points :
- regular sample drain (no output sealing)
(output side on the left side) sufficient to
ensure a regular drain.
Содержание SODIMAT 9073
Страница 1: ...SODIMAT Model 9073 Instruction manual 03 2001 621 190 073 Rev D ...
Страница 2: ......
Страница 12: ......
Страница 20: ......
Страница 27: ...SODIMAT Model 9073 3 7 3 3 PREAMPLIFIER Figure 3 4 Preamplifier connections ...
Страница 34: ......
Страница 36: ...SODIMAT Model 9073 5 2 5 1 General configuration and calibrations ...
Страница 38: ...SODIMAT Model 9073 5 4 5 2 Alarm configuration ...
Страница 40: ...SODIMAT Model 9073 5 6 5 3 Analogue output configuration ...
Страница 42: ...3 2 1 0 0 1 1 0 2 3 4 5 SODIMAT Model 9073 5 8 5 4 Serial interface configuration ...
Страница 56: ......
Страница 58: ......
Страница 65: ...SODIMAT Model 9073 10 4 Figure 10 2 Mixed bed for AUTOCAL version 1 Mixed bed cartridge 2 Control valve ...
Страница 69: ...SODIMAT Model 9073 10 8 Figure 10 4 Changing the pump tubing ...
Страница 74: ......
Страница 80: ......