SODIMAT - Model 9073
5-5
Two different ways to acquit manually the alarm system :
1. By pressing ENTER.
2. Remotely by the single pole contact (see figure 3-2)
Command
Function
Option
500
0
Limits S1 and S2 not active
1
Limit S1 active and limit S2 not active
2*
Limits S1 and S2 active
505
0*
S1 defined as low limit setpoint
1
S1 defined as high limit setpoint
506
[0.01*]
S1 value in ppb or ppm
507
[0*]
Delay time between 0...9999 seconds
515
0
S2 defined as low limit setpoint
1*
S2 defined as high limit setpoint
516
[999.9*]
S2 value in ppb or ppm
517
[0*]
Delay time between 0...9999 seconds
591
0*
Relay S1 normally open if no alarm condition
1
Relay S1 activated if no alarm condition
592
0*
Relay S2 normally open if no alarm condition
1
Relay S2 activated if no alarm condition
593
0
Relay S3 off
1*
Relay S3 activated by an alarm system with manual reset
2
Relay S3 activated by an alarm system with automatic reset
Содержание SODIMAT 9073
Страница 1: ...SODIMAT Model 9073 Instruction manual 03 2001 621 190 073 Rev D ...
Страница 2: ......
Страница 12: ......
Страница 20: ......
Страница 27: ...SODIMAT Model 9073 3 7 3 3 PREAMPLIFIER Figure 3 4 Preamplifier connections ...
Страница 34: ......
Страница 36: ...SODIMAT Model 9073 5 2 5 1 General configuration and calibrations ...
Страница 38: ...SODIMAT Model 9073 5 4 5 2 Alarm configuration ...
Страница 40: ...SODIMAT Model 9073 5 6 5 3 Analogue output configuration ...
Страница 42: ...3 2 1 0 0 1 1 0 2 3 4 5 SODIMAT Model 9073 5 8 5 4 Serial interface configuration ...
Страница 56: ......
Страница 58: ......
Страница 65: ...SODIMAT Model 9073 10 4 Figure 10 2 Mixed bed for AUTOCAL version 1 Mixed bed cartridge 2 Control valve ...
Страница 69: ...SODIMAT Model 9073 10 8 Figure 10 4 Changing the pump tubing ...
Страница 74: ......
Страница 80: ......