Version 2.0
24
6_26931B ELS2100_2100 Ice Se rvice Manual_vsn2.0_maste r.doc
Replacing The Solvent Leak Assembly
When required:
Faulty Solvent leak sensor assembly
Tools required:
Allen Keys
Spanners
Screwdriver
Parts required:
Leak Sensor Assembly (# 0860-0520)
1) Open the detector’s side panel
using the access key, as shown.
2) Ensure there is no liquid in the
leak sensor well (located
underneath the nebuliser chamber).
Dry with tissue if necessary.
Remove the two screws on the
leak sensor assembly
3) Disconnect the cable that
connects the leak sensor assembly
to the main control board.
Replace the leak sensor assembly
and reconnect the cable.
If necessary replace the leak sensor
cable assembly.