17
DMC Repair Manual
Parts Replacement
Cautions
Electrostatic
-
Digital Microscope Camera circuitry can easily be damaged by small
Discharge
static discharges. Always use an anti-static mat and an ESD wrist-
Warning
strap when disassembling, assembling or repairing the camera with
the bottom extrusion removed.
Inspection and -
Whenever disassembling a Digital Microscope Camera for parts
Cleaning
replacement, inspect it for dust and debris. If necessary, clean the
camera to assure that dust in the optical path does not degrade its
performance.
Connectors
-
The connectors in the Digital Microscope Camera have locking tabs
(Figure 2-1) which secures the flex cable in place. To remove a flex
cable, first pull back the two (2) locking tabs and pull the flex cable free
of the connector. To install a flex cable, align it in the connector and
make sure it is fi\ully seated befor pushing in the two (2) locking tabs.
Figure 2-1. Typical flex connector
Screw
-
Do not reuse the front and rear panel screws. Always use new
Replacement
screws (PH PA 4-40 x .75) when reinstalling the front and rear
(Front /Rear
panels. These screws are coated to prevent them from backing
Panels)
out.
Locking Tab
Locking Tab
Flex Cable
Connector