PME miniDOT Скачать руководство пользователя страница 9

Precision Measurement Engineering, Inc.
(760) 727-0300
www.pme.com

9

If the marine growth remains, then use Q-tips to gently wipe the sensing foil after it has been
softened by soaking in vinegar or perhaps diluted HCl. After cleaning the sensing foil, then it
should be rinsed well in clean tap water before storing or reuse.  Do not use other organic
solvents such as acetone, chloroform, and toluene since these and others will damage the
sensing foil.

The sensing foil can also be cleaned using a 3% H2O2 solution or rinsing it with ethanol.

The white pressure housing and black end cap can be gently scrubbed.

AA Alkaline Battery Life

 – Alkaline batteries will give somewhat less performance than

lithium, especially at low temperatures.  Alkaline batteries are superior to lithium in one way:
you can determine how much battery life remains by measuring the battery terminal voltage.
For short deployments of a month or two, then alkaline batteries will provide adequate
performance.  For longer deployments, or for deployments in cold environments, then
substitute lithium batteries.

AA Lithium Battery Life

 – The miniDOT Logger consumes battery power mostly from the

measurement of dissolved oxygen, but also slightly from simply keeping track of time, writing
files, sleeping, and other activities.  The following table shows the approximate endurance of
the miniDOT Logger when powered by the Energizer L91 AA lithium / ferrous disulfide
batteries:

SAMPLE INTERVAL

(minutes)

Main AA Battery Life

(months)

Number of Samples

1

12

500K

10

>12

>52,000

60

>12

>8,000

Keep a general record of the miniDOT Logger’s number of samples.  It is not possible to
accurately tell the charge state of a lithium battery by measuring its terminal voltage.  If you
have a general idea of the number of samples already obtained on a battery, then you can
make a guess as to how many more samples remain.

The numbers in the table above are, at the time of this writing, based upon extrapolations of
testing of 500K samples acquired at a 5-second interval.  The 1-year performance at 1
minute is very likely.  Performance at longer sample intervals will be much longer, but how
long is difficult to predict.   In any event, these AA batteries are easily available and relatively
inexpensive compared to the cost of the miniDOT Logger.  PME suggests you replace the
batteries often, especially before any long (months) measurement deployments.

Monitor battery voltage in the miniDOTControl program.  You cannot tell from terminal
voltage of a lithium battery how long the battery will last, but you can tell if it will die soon.

Содержание miniDOT

Страница 1: ...Precision Measurement Engineering Inc 760 727 0300 www pme com 1 Precision Measurement Engineering Inc www pme com User s Manual 2018...

Страница 2: ...iginal manufacturer PME makes no warranty either expressed or implied that the sensors will be operable after they are exposed to adverse environmental conditions such as bio fouling oil fouling freez...

Страница 3: ...14JUL16 Added Storage When Not in Use to sec 1 2 2014SEP24 Added pictures reworded Removed wire mesh installation instructions 2014NOV14 Added update Java instructions Removed statement saying alkalin...

Страница 4: ...Details 1 3 Overview General Maintenance Chapter 2 Software 2 1 Overview and Software Installation 2 2 miniDOTControl 2 3 miniDOTPlot 2 4 miniDOTConcatenate Chapter 3 miniDOT Logger 3 1 Overview 3 2 O...

Страница 5: ...ng the time the measurements were made are recorded in text files in the folder having the serial number of your miniDOT Logger These files can be copied onto any Windows or Mac HOST computer This Man...

Страница 6: ...uter 6 Suggested but optional Delete the measurement folder but NOT the miniDOTControl program or the other jar programs 7 Optional Run the miniDOTControl program to see the state of the miniDOT Logge...

Страница 7: ...le based on the miniDOT Logger s internal clock and expressed in YYYY MM DD HHMMSSZ txt format For example a file having the first measurement on September 9 2014 at 17 39 00 UTC will be named 2014 09...

Страница 8: ...miniDOT Logger s green connection light come on and your Android device will display a message about preparing USB storage When USB storage ready is shown then use your file explorer program to move y...

Страница 9: ...xygen but also slightly from simply keeping track of time writing files sleeping and other activities The following table shows the approximate endurance of the miniDOT Logger when powered by the Ener...

Страница 10: ...ers in type or charge level from the other and the miniDOT Logger runs them to full discharge then one battery may leak SEE SECTION 3 4 FOR CAUTION ON BATTERY PLACEMENT Err on the side of caution when...

Страница 11: ...white pressure housing is screwed back on to the black end cap then these drops can become trapped inside the miniDOT Logger Be sure to carefully dry the o ring and adjacent surfaces especially under...

Страница 12: ...s temperature to come to equilibrium with the water During the experiment find the local air pressure either from measurements or from a local weather station Watch out weather stations often report b...

Страница 13: ...stainless steel screws in the black end cap There are no user serviceable parts here If the screws are removed then you will damage the miniDOT Logger and it will have to be returned for repair Stora...

Страница 14: ...lready be installed on the HOST computer You can test this by running the miniDOTPlot program If this program displays its graphical user interface then the JRE is installed If not then the JRE can be...

Страница 15: ...passed since time was last set then the miniDOT Logger s clock will be set and the checkmark icon will appear If the HOST computer is not connected to the Internet then no time services will occur The...

Страница 16: ...that contains the files recorded by the miniDOT Logger If the miniDOTPlot program is run directly from the miniDOT Logger then the program will suggest the folder that is located on the miniDOT Logge...

Страница 17: ...simply reads the miniDOT Logger s measurement files again End the miniDOTPlot program by closing the window Special note plotting of sample sets of more than 200K samples may consume all the memory a...

Страница 18: ...to the HOST computer and select them there The file access to the miniDOT Logger is slow The miniDOT Logger measurement folders must NOT contain any files besides those the miniDOT Logger recorded and...

Страница 19: ...inum chassis Try not to touch the circuit board When closing the miniDOT Logger inspect the o ring and the interior of the white pressure housing for debris Lubricate the o ring and screw the white pr...

Страница 20: ...is controlled by the Logger Control Switch position The switch position may be changed while the USB is connected NOTE miniDOT units equipped with a 16GB SD card will take longer to appear as a useab...

Страница 21: ...you complete the battery installation At this time the logger will enter the mode selected by the Logger Control Switch which should initially be Halt from Step 1 Please be aware that the warranty wi...

Страница 22: ...ntly tighten CAUTION Environments where debris can get trapped inside the sensing area should be avoided when using this product PME recommends using the Cu plate in such environments How to install t...

Страница 23: ...0 www pme com 23 2 Position the Copper Plate faced down so that the notches in the Copper Plate perfectly align over sensing foil and over the screw holes 4 Install the three pan head screws included...

Отзывы: