Precision Measurement Engineering, Inc.
(760) 727-0300
www.pme.com

19

3.2  Opening and Closing the miniDOT Logger

The miniDOT Logger’s circuitry is contained in a waterproof housing that must be opened.
Unscrewing the white pressure housing from the black end cap opens the miniDOT Logger.
This is similar to opening a flashlight.  Turn the white pressure housing counter clockwise
relative to the black end cap.   Close the miniDOT Logger by reversing this procedure after
being sure that the o-ring is free from debris.  Lubricate the o-ring occasionally with silicone
oil intended for buna-N o-ring material.

Please attempt to handle the miniDOT Logger by only touching the aluminum chassis. Try
not to touch the circuit board.

When closing the miniDOT Logger, inspect the o-ring and the interior of the white pressure
housing for debris. Lubricate the o-ring, and screw the white pressure housing onto the black
end cap until the white pressure housing just touches the black end cap.  Do not tighten!  The
miniDOT Logger tends to get a little tighter during deployment.

If you cannot open the miniDOT Logger by yourself, then find another person with strong
hands.  This person should grip the black end cap while the other person turns the white
pressure housing.

Caution:  Do NOT remove the stainless screws in the black end cap.  If this is done, then the
miniDOT Logger will be permanently damaged and must be returned for repair.

3.3  Electrical Connections and Controls

Removal of the cover reveals the miniDOT Logger’s connections and controls, shown below.

Содержание miniDOT

Страница 1: ...Precision Measurement Engineering Inc 760 727 0300 www pme com 1 Precision Measurement Engineering Inc www pme com User s Manual 2018...

Страница 2: ...iginal manufacturer PME makes no warranty either expressed or implied that the sensors will be operable after they are exposed to adverse environmental conditions such as bio fouling oil fouling freez...

Страница 3: ...14JUL16 Added Storage When Not in Use to sec 1 2 2014SEP24 Added pictures reworded Removed wire mesh installation instructions 2014NOV14 Added update Java instructions Removed statement saying alkalin...

Страница 4: ...Details 1 3 Overview General Maintenance Chapter 2 Software 2 1 Overview and Software Installation 2 2 miniDOTControl 2 3 miniDOTPlot 2 4 miniDOTConcatenate Chapter 3 miniDOT Logger 3 1 Overview 3 2 O...

Страница 5: ...ng the time the measurements were made are recorded in text files in the folder having the serial number of your miniDOT Logger These files can be copied onto any Windows or Mac HOST computer This Man...

Страница 6: ...uter 6 Suggested but optional Delete the measurement folder but NOT the miniDOTControl program or the other jar programs 7 Optional Run the miniDOTControl program to see the state of the miniDOT Logge...

Страница 7: ...le based on the miniDOT Logger s internal clock and expressed in YYYY MM DD HHMMSSZ txt format For example a file having the first measurement on September 9 2014 at 17 39 00 UTC will be named 2014 09...

Страница 8: ...miniDOT Logger s green connection light come on and your Android device will display a message about preparing USB storage When USB storage ready is shown then use your file explorer program to move y...

Страница 9: ...xygen but also slightly from simply keeping track of time writing files sleeping and other activities The following table shows the approximate endurance of the miniDOT Logger when powered by the Ener...

Страница 10: ...ers in type or charge level from the other and the miniDOT Logger runs them to full discharge then one battery may leak SEE SECTION 3 4 FOR CAUTION ON BATTERY PLACEMENT Err on the side of caution when...

Страница 11: ...white pressure housing is screwed back on to the black end cap then these drops can become trapped inside the miniDOT Logger Be sure to carefully dry the o ring and adjacent surfaces especially under...

Страница 12: ...s temperature to come to equilibrium with the water During the experiment find the local air pressure either from measurements or from a local weather station Watch out weather stations often report b...

Страница 13: ...stainless steel screws in the black end cap There are no user serviceable parts here If the screws are removed then you will damage the miniDOT Logger and it will have to be returned for repair Stora...

Страница 14: ...lready be installed on the HOST computer You can test this by running the miniDOTPlot program If this program displays its graphical user interface then the JRE is installed If not then the JRE can be...

Страница 15: ...passed since time was last set then the miniDOT Logger s clock will be set and the checkmark icon will appear If the HOST computer is not connected to the Internet then no time services will occur The...

Страница 16: ...that contains the files recorded by the miniDOT Logger If the miniDOTPlot program is run directly from the miniDOT Logger then the program will suggest the folder that is located on the miniDOT Logge...

Страница 17: ...simply reads the miniDOT Logger s measurement files again End the miniDOTPlot program by closing the window Special note plotting of sample sets of more than 200K samples may consume all the memory a...

Страница 18: ...to the HOST computer and select them there The file access to the miniDOT Logger is slow The miniDOT Logger measurement folders must NOT contain any files besides those the miniDOT Logger recorded and...

Страница 19: ...inum chassis Try not to touch the circuit board When closing the miniDOT Logger inspect the o ring and the interior of the white pressure housing for debris Lubricate the o ring and screw the white pr...

Страница 20: ...is controlled by the Logger Control Switch position The switch position may be changed while the USB is connected NOTE miniDOT units equipped with a 16GB SD card will take longer to appear as a useab...

Страница 21: ...you complete the battery installation At this time the logger will enter the mode selected by the Logger Control Switch which should initially be Halt from Step 1 Please be aware that the warranty wi...

Страница 22: ...ntly tighten CAUTION Environments where debris can get trapped inside the sensing area should be avoided when using this product PME recommends using the Cu plate in such environments How to install t...

Страница 23: ...0 www pme com 23 2 Position the Copper Plate faced down so that the notches in the Copper Plate perfectly align over sensing foil and over the screw holes 4 Install the three pan head screws included...