- 55 -
Place the Cephalostat calibrating tool (hatchet) into the patient’s right
ear post holder with the point going towards the back or column.
(Figure 99)
Figure 99
NOTE: Change binnings by clicking Settings and select Binnings then the appropriate binning in
Dimax3tool. (Figure 100)
Figure 100
5.2
Computer
Click on Start, Click on All Programs then Planmeca then Dimax3 Tool.