- 50 -
3.1
Left / Right
If the beam is too left or right, loosen the two screws closest to the tubehead
with a 4mm and slide left or right depending on what is needed. (Figure 89)
Figure 89
Take another exposure and check alignment, if still off adjust again. If aligned
then re-tighten the screws and take a final exposure to make sure nothing has
shifted.
3.2
High / Low
If the beam is too high or low, loosen the three screws on the back of the
beaver tall that are at the top closest the sensor holder. (Figure 90)
Figure 90