- 52 -
3.3.b Right
Un-tighten the top side screw with a 2.5mm allen and tighten the
bottom screw in. (See Figure 92)
Take an exposure, adjust as needed.
4
Beam Check
NOTE: Remember to put the sensor into the sensor holder before continuing.
NOTE: Make sure that the x-ray machine is connected to computer
4.1 Setting up beam check
Press the
i
icon on the lower hand corner of the touch screen. (Figure 93)
Figure 93
Press
Beam Check.
(Figure 94)
Figure 94
NOTE: If the sensor was calibrated first, remove the red calibration block from the collimator.