- 51 -
NOTE: If unable to move beaver tall up or down after loosen these screws, take a 2.5mm allen and loosen
slightly the two screws on the beaver tall that are towards the front of the machine. (Figure 91)
Figure 91
Re-tighten the middle screw to hold the adjusted position and take another
exposure, if still off adjust again. If aligned then re-tighten the other two
screws and the two side screws, if they were un-tightened. Take another to
check that nothing has shifted.
3.3
Angled
If the beam is angled, loosen the two side screws that are facing the sensor
holder with a 4mm allen. (See Figure 91)
3.3.a Left
Un-tighten the bottom side screw with a 2.5mm allen and tighten the
top screw in. (Figure 92)
Figure 92
Take an exposure, adjust as needed.