The beam image should appear within the borders of the rectangle markings on the
fluorescent screen. (Figure 12)
U P
D O W N
Figure 12
If not too left, right, up, down, or angled see below.
1.1 High / Low
If the beam is too high or too low, loosen up the two screws at the top and
bottom positions on the collimator with a 3mm allen. (Figure 13)
Figure 13
Take another exposure until the beam is aligned.