- 23 -
Press all the arrows at the bottom of the touch screen, turning each
from white to red. (Figure 27)
Figure 27
1.2.b ON
Press the smile in the middle right of the right of the touch screen. (See
Figure 27) Then Press all the arrows at the bottom of the touch screen,
turning each from red to white.
NOTE: Make sure the red calibration block is placed onto the collimator before calibrating. (Figure 28)
Figure 28