3. Pull out the coolant tank.
4. Remove the coolant tank lid.
5. Fill or empty the coolant tank. The capacity of the coolant tank is 1.8
litres.
6. Close the coolant tank lid.
7. Make sure that the feed line and the float are in the uppermost position.
8. Push the coolant tank all the way to the back of the unit.
9. Push the feed line and the float downwards.
10. Make sure that the feed line and the float are positioned in the recess
provided in the lid of the coolant tank and that the function of the float is
guaranteed.
11. Close the coolant tank access door.
11.4 Maintenance plan
11.4.1 Before every start-up of unit
•
Clean the workpiece holder
•
Clean the tool holder
•
Clean the tool shank on all tools
•
Clean the length probe
•
Clean the interior of the unit
11.4.2 Daily
•
Check the water separator and empty if necessary
11.4.3 Weekly or after tool breakage
•
Remove the collet chuck
•
Clean the collet chuck
•
Check the collet chuck for damage
11 Preventive maintenance