4. Open the access door to the coolant tank by pulling the door corners.
5. Slide the suction device fully upwards.
6. Remove the coolant tank.
7. Thoroughly clean all filters and the sump of contaminants.
8. Fill the sump with 1.8 L of diluted cooling liquid mixture (observe the
mixture ratio according to the coolant system description).
9. Reinsert the cooling liquid tank.
10. Slide the suction device fully downwards again, see step 5.
11. Close the access door to the coolant tank, see step 4.
12. Start the milling with the coolant system switched on and without an
extraction system.
8.8 Modes of operation
The unit can be operated in three modes.
8 Operating milling unit
User's manual
Planmeca PlanMill 50 S 51