11 Preventive maintenance
11.1 Safety instructions for maintenance
Regular maintenance and care will increase the operational reliability and
service life of the unit. Turn the unit off before any cleaning or maintenance
work for which a particular position (
Service position or Cleaning position) is
necessary. Remove the power supply plug to prevent unintentional
reconnection.
WARNING
Mortal danger! Parts of the electrical system remain under power even when
the main switch is off. Please wait a few minutes before beginning
maintenance work.
To clean the unit use a damp cloth and, if necessary, a suitable cleaning
agent. Never use abrasive cleaning agents or detergents that may attack
plastics, rubber or paint.
These may damage hoses (pneumatic), cables and seals. Cleaning agents
may contain substances that are hazardous to health.
Comply with the manufacturer’s instructions.
WARNING
Mortal danger! All work on the electrical system, such as connection to the
network, maintenance and repairs should only be performed by qualified
personnel.
Parts of the electrical system remain under power even when the main
switch is off.
NOTE
All work on the pneumatic system such as commissioning, maintenance and
repairs may only be performed by qualified personnel.
Before working on the pneumatic system the system must be shut down and
depressurised.
11.2 Replacement parts
WARNING
Risk of injury due to using incorrect replacement parts! Using incorrect or
substandard replacement parts can pose a risk for personnel and potentially
cause damage, incorrect operation or total failure of the unit.
•
Use only the manufacturer’s original replacement parts or replacement
parts authorised by the manufacturer.
•
In case of doubt contact the local dealer.
NOTE
The use of unapproved replacement parts will void the manufacturer’s
warranty.