2. Remove the clean-water bottle.
First, turn the Water bottle switch toward 'OFF'. Then, take a hold of the
bottom of the bottle, turn the bottle slightly so that the pin moves along
the groove and pull the bottle downwards.
3. Rinse the clean-water bottle and fill it with clean water.
4. Place the clean-water bottle back in its place in the dental unit and close
the unit door.
5. Start the short flushing cycle.
Balanced instrument arms: Start the flushing cycle by bending the
instrument arms of those instruments (including syringe) that are to be
flushed to an angle of at least 90° and simultaneously pressing
Instrument spray (you will hear a signal tone) and holding it for 3
seconds. Release the button and arms when you hear a second signal
tone.
Hanging-tube instruments: Start the flushing cycle by pressing
Instrument spray (you will hear a signal tone) and holding it for 3
seconds. Release the button when you hear a second signal tone.
The system will first identify the instruments and then each instrument
hose is flushed with air and water for 30 seconds in its turn.
25 Clean water system
User's manual
Planmeca Compact i 135
Содержание Compact i Classic v2